Performance Specifications Appendix<br />
ESU Protection: 3 Lead and 5 Lead ECG shall meets standard<br />
IEC 601-2-25, clause 36.202.7 for<br />
functionality following ESU energy exposure.<br />
The 12 Lead ECG function provides no ESU<br />
protection.<br />
Electro Surgical Unit Noise Suppression: Peak noise is less than ±<br />
2mV from ECG baseline when used with<br />
AAMI compatible cables. Noise level is<br />
unspecified when the monitor is used with HP<br />
cables. The 12 Lead ECG function provides<br />
no ESU protection or noise suppression.<br />
ESU Withstand: 3 Lead and 5 Lead ECG shall withstand ESU<br />
stress from a High Frequency Surgical Units<br />
operating at 300 Watts in cut mode and 100<br />
Watts on coagulate mode. The 12 Lead ECG<br />
function provides no ESU protection.<br />
Input Bias Current: < 100 nA per lead, excluding driven lead<br />
(ANSI/AAMI EC13-1992, 3.2.5 or ANSI/<br />
AAMI EC 11-1991 3.2.10).<br />
Input Signal Range: ± 5.0 mV, minimum<br />
DC Offset: ± 300 mV minimum (ANSI/AAMI EC13-<br />
1992, 3.2.9.1 or ANSI/AAMI EC 11-1991,<br />
3.2.3).<br />
Input Impedance: > 2.5 MΩ single ended at 10Hz per ANSI/<br />
AAMI EC13-1992, 3.2.9.2 or ANSI/AAMI<br />
EC11-1991, 3.2.9.<br />
Noise: Less than 30 µV p-p, referred to input, through<br />
shielded 51KΩ resistors in parallel with 47nF<br />
capacitors (ANSI/AAMI EC13-1992, 3.2.9.3<br />
or ANSI/AAMI EC 11-1991, 3.2.12.1).<br />
Overall System Error: 5% or 40uV, whichever is greater (ANSI/<br />
AAMI EC-13-1992, 3.2.9.8 or ANSI/AAMI<br />
EC 11-1991, 3.2.7.1)<br />
Gain Stability: Per ANSI/AAMI EC13-1992, 3.2.9.5 (d) or<br />
ANSI/AAMI EC 11-1991, 3.2.4.4.<br />
6 - 4 0070-10-0649-01 <strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions