Status Messages Operation<br />
3.22.2 SpO 2 Messages<br />
Always have an alternate method of BP verification available.<br />
On vasoconstricted patients, failure to evacuate air from the cuff can distort BP<br />
measurement.<br />
Do not place cuff on extremity that has an IV.<br />
Cuff should be at heart level.<br />
The following messages pertain to SpO 2 Operation.<br />
MESSAGE REASON ACTION<br />
SpO2: No Sensor Sensor is not plugged in to the<br />
<strong>Passport</strong> 2.<br />
Plug the sensor into the monitor<br />
SpO 2: Sensor Off (Masimo<br />
Only)<br />
Sensor may not be connected<br />
to the patient.<br />
SpO 2: Interference Noise detected on the pulse<br />
signal prevents pulse<br />
discrimination.<br />
SpO 2: Pulse Search Hardware settings are being<br />
adjusted in order to<br />
discriminate a pulse waveform.<br />
SpO2: No Pulse (Nellcor Only) No detectable pulse is<br />
measured.<br />
SpO2: Failure The system has detected an<br />
unrecoverable failure of the<br />
SpO2 system.<br />
SpO2: Low Perfusion<br />
(Masimo Only)<br />
SpO2: Too Much Light<br />
(Masimo Only)<br />
SpO 2 : Unrecognized Sensor<br />
(Masimo Only)<br />
Check patient connection.<br />
Decrease patient motion, check<br />
sensor.<br />
Change to site where pulse is<br />
stronger if patient is<br />
vasoconstricted. Change or<br />
readjust sensor if loose.<br />
Check to patient connection<br />
and patient status.<br />
Power cycle unit. If message<br />
reappears, call <strong>Mindray</strong> DS<br />
Service.<br />
Patient perfusion is low. Check to patient connection<br />
and patient status.<br />
There is too much ambient<br />
room light for the sensor to<br />
function properly.<br />
The sensor is not recognized<br />
by the Monitor.<br />
SpO2: Communication Error The monitor and the SpO2 modules are not<br />
communicating properly.<br />
SpO2: Board Fault Masimo board failed to<br />
operate properly.<br />
Minimize the room light<br />
around the patient. Check<br />
sensor.<br />
Replace the sensor with a<br />
<strong>Mindray</strong> DS recommended<br />
sensor.<br />
Power the unit OFF / ON. If<br />
problem persists, contact<br />
<strong>Mindray</strong> DS Service.<br />
Contact <strong>Mindray</strong> DS Customer<br />
Support.<br />
SpO2: Sensor Fault Defective Sensor. Replace sensor.<br />
SpO2: Motion (Nellcor Only) Motion is detected. Decrease patient motion, check<br />
sensor.<br />
SpO 2: Check Sensor (Nellcor<br />
Only)<br />
The SpO 2 module has sensed<br />
a poor connection or a bad<br />
sensor.<br />
Reconnect the same sensor. If<br />
problem persists, replace<br />
sensor.<br />
3 - 80 0070-10-0649-01 <strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions