Status Messages Operation<br />

3.22.2 SpO 2 Messages<br />

Always have an alternate method of BP verification available.<br />

On vasoconstricted patients, failure to evacuate air from the cuff can distort BP<br />

measurement.<br />

Do not place cuff on extremity that has an IV.<br />

Cuff should be at heart level.<br />

The following messages pertain to SpO 2 Operation.<br />

MESSAGE REASON ACTION<br />

SpO2: No Sensor Sensor is not plugged in to the<br />

<strong>Passport</strong> 2.<br />

Plug the sensor into the monitor<br />

SpO 2: Sensor Off (Masimo<br />

Only)<br />

Sensor may not be connected<br />

to the patient.<br />

SpO 2: Interference Noise detected on the pulse<br />

signal prevents pulse<br />

discrimination.<br />

SpO 2: Pulse Search Hardware settings are being<br />

adjusted in order to<br />

discriminate a pulse waveform.<br />

SpO2: No Pulse (Nellcor Only) No detectable pulse is<br />

measured.<br />

SpO2: Failure The system has detected an<br />

unrecoverable failure of the<br />

SpO2 system.<br />

SpO2: Low Perfusion<br />

(Masimo Only)<br />

SpO2: Too Much Light<br />

(Masimo Only)<br />

SpO 2 : Unrecognized Sensor<br />

(Masimo Only)<br />

Check patient connection.<br />

Decrease patient motion, check<br />

sensor.<br />

Change to site where pulse is<br />

stronger if patient is<br />

vasoconstricted. Change or<br />

readjust sensor if loose.<br />

Check to patient connection<br />

and patient status.<br />

Power cycle unit. If message<br />

reappears, call <strong>Mindray</strong> DS<br />

Service.<br />

Patient perfusion is low. Check to patient connection<br />

and patient status.<br />

There is too much ambient<br />

room light for the sensor to<br />

function properly.<br />

The sensor is not recognized<br />

by the Monitor.<br />

SpO2: Communication Error The monitor and the SpO2 modules are not<br />

communicating properly.<br />

SpO2: Board Fault Masimo board failed to<br />

operate properly.<br />

Minimize the room light<br />

around the patient. Check<br />

sensor.<br />

Replace the sensor with a<br />

<strong>Mindray</strong> DS recommended<br />

sensor.<br />

Power the unit OFF / ON. If<br />

problem persists, contact<br />

<strong>Mindray</strong> DS Service.<br />

Contact <strong>Mindray</strong> DS Customer<br />

Support.<br />

SpO2: Sensor Fault Defective Sensor. Replace sensor.<br />

SpO2: Motion (Nellcor Only) Motion is detected. Decrease patient motion, check<br />

sensor.<br />

SpO 2: Check Sensor (Nellcor<br />

Only)<br />

The SpO 2 module has sensed<br />

a poor connection or a bad<br />

sensor.<br />

Reconnect the same sensor. If<br />

problem persists, replace<br />

sensor.<br />