Operation Status Messages<br />
MESSAGE REASON ACTION<br />
GM: Agent<br />
Mismatch - ENF<br />
GM: Agent<br />
Mismatch - SEV<br />
GM: Agent<br />
Mismatch - DES<br />
GM: Unknown<br />
Agent<br />
GM: Cannot Zero...<br />
RETRYING<br />
Appears when the system detects<br />
Enflurane as the primary agent and the<br />
manually selected agent is not Enflurane.<br />
Appears when the system detects<br />
Sevoflurane as the primary agent and the<br />
manually selected agent is not<br />
Sevoflurane.<br />
Appears when the system detects<br />
Desflurane as the primary agent and the<br />
manually selected agent is not<br />
Desflurane.<br />
Appears when the system detects a gas<br />
that does not match the spectroscopic<br />
signatures of the five known anesthetic<br />
agents.<br />
Appears when the <strong>Passport</strong> requests a<br />
Zeroing (either on the automatic cycle or<br />
by a user request), and the Gas Module<br />
is unable to initialize the cycle.<br />
GM: ET CO 2 High Appears when the End Tidal CO 2<br />
measurement is greater than or equal to<br />
the value set for the ET CO 2 High Alarm.<br />
GM: ET CO 2 Low Appears when the End Tidal CO 2<br />
measurement is less than or equal to the<br />
value set for the ET CO 2 Low Alarm.<br />
GM: INSP CO 2<br />
High<br />
Appears when the FiCO 2 measurement is<br />
greater than or equal to the value set for<br />
the INSP CO 2 High Alarm.<br />
GM: ET O 2 High Appears when the End Tidal O 2<br />
measurement is greater than or equal to<br />
the value set for the ET O 2 High Alarm.<br />
GM: ET O 2 Low Appears when the End Tidal O 2<br />
measurement is less than or equal to the<br />
value set for the ET O 2 Low Alarm.<br />
GM: INSP O 2 High Appears when the FiO 2 measurement is<br />
greater than or equal to the value set for<br />
the INSP O 2 High Alarm.<br />
GM: INSP O 2 Low Appears when the FiO 2 measurement is<br />
less than or equal to the value set for the<br />
INSP O 2 Low Alarm.<br />
GM: INSP N 2O<br />
High<br />
Appears when the Inspired N 2O<br />
measurement is greater than or equal to<br />
the value set for the INSP N 2O High<br />
Alarm.<br />
GM: ET Agent High Appears when the End Tidal Anesthetic<br />
Agent measurement is greater than or<br />
equal to the value set for the ET Agent<br />
High Alarm.<br />
GM: ET Agent Low Appears when the End Tidal Anesthetic<br />
Agent measurement is less than or equal<br />
to the value set for the ET Agent Low<br />
Alarm.<br />
Match the Agent administered<br />
with the Agent selected, or<br />
select Agent Auto ID.<br />
Match the Agent administered<br />
with the Agent selected, or<br />
select Agent Auto ID.<br />
Match the Agent administered<br />
with the Agent selected, or<br />
select Agent Auto ID.<br />
Use recognized agent.<br />
Allow system to retry without<br />
intervention. If problem persists<br />
call for service.<br />
Check patient.<br />
Check patient.<br />
Check patient.<br />
Check patient.<br />
Check patient.<br />
Check patient.<br />
Check patient.<br />
Check patient.<br />
Check patient.<br />
Check patient.<br />
<strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 3 - 83