Operation Non-Invasive Blood Pressure Measurements (NIBP)<br />
2. Attach cuff hose to NIBP Connector.<br />
3. Apply the cuff to the patient. To reduce errors, the cuff should be fitted snugly, with little<br />
or no air present within the cuff. Be sure the cuff lies directly against the patient’s skin.<br />
No clothing should come between the patient and the cuff.<br />
NOTE: The NIBP cuff should not be placed on a limb that is being<br />
utilized for any other medical procedure. For example, an IV<br />
catheter or an SpO 2 sensor.<br />
4. If not already selected, select the Patient Size through the Patient Menu as described in<br />
Section 2.3. Choices are Adult, Pediatric or Neonate.<br />
5. If necessary, change the initial cuff inflation pressure through the NIBP Menu.<br />
Initial cuff inflation pressures depend on the patient size setting. The choices of cuff inflation<br />
are:<br />
PATIENT SIZE<br />
SETTING<br />
INITIAL CUFF<br />
INFLATION VALUES<br />
6. Press START to begin an NIBP measurement.<br />
DEFAULT<br />
SETTING<br />
MAXIMUM<br />
INFLATION VALUES<br />
Adult 100 - 280 mmHg 180 mmHg 300 mmHg<br />
Pediatric 60 - 180 mmHg 140 mmHg 200 mmHg<br />
Neonate 40 - 120 mmHg 100 mmHg 150 mmHg<br />
NOTE: Inflate the cuff only after proper application to the patient’s<br />
limb. Cuff damage can result if the cuff is left unwrapped<br />
and then inflated.<br />
The cuff begins to inflate to the selected cuff pressure. After reaching the selected value the<br />
cuff begins to slowly deflate and the <strong>Mindray</strong> DS <strong>Passport</strong> 2/<strong>Passport</strong> 2 LT Monitor<br />
collects oscillometric pulsations.<br />
If the initial cuff inflation is found to be inadequate, the unit retries with a higher inflation<br />
pressure (+50 mmHg in the adult mode; +40 mmHg in the pediatric and neonate modes).<br />
Have the patient remain still to avoid the introduction of unnecessary motion artifact. After<br />
the cuff pressure drops below the diastolic pressure, the results of the measurement are<br />
displayed.<br />
If NIBP is the only parameter being measured with the <strong>Passport</strong> 2/<strong>Passport</strong> 2 LT, a heart<br />
rate can be derived from NIBP. The HR source menu selection must be in the Auto mode (i.e.,<br />
not selected for ECG, IBP or SpO2 ) with no heart rate alarm limits set. (See Alarms<br />
Section 3.19 for details). If another heart rate source is available, the NIBP heart rate will be<br />
replaced by the heart rate from the selected source.<br />
<strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 3 - 9