Decontamination of Monitor User Maintenance<br />
4.3 Decontamination of Monitor<br />
WARNING: Perform the decontamination process with the unit powered<br />
down and power cord removed.<br />
Decontamination of a unit that has come in contact with a biological material can be<br />
performed using LpH SE Germicidal detergent. Apply a small amount of detergent to a<br />
disposable wipe (paper based) and wipe down the outside of the unit. Discard the wipe<br />
appropriately. After waiting 10 minutes, use a clean dry wipe to dry the unit.<br />
CAUTION: During the decontamination process, do not get the LpH SE<br />
Germicidal detergent into any vent openings.<br />
4.4 Care and Cleaning of SpO 2 Sensors<br />
NOTE: Refer to the individual instruction sheets that are packaged<br />
with each sensor.<br />
Daily, check the sensors and cables for signs of damage. Replace as required.<br />
The sensors should be cleaned before and after each patient’s use.<br />
Wipe the patient contact area using a soft cloth with mild soap and water solution or<br />
isopropyl alcohol. Hydrogen peroxide can be used to remove dried blood on all<br />
accessible surfaces.<br />
Let the sensor completely dry before using.<br />
CAUTION: When cleaning SpO 2 sensors, do not use excessive amounts<br />
of liquid. Wipe the sensor surface with a soft cloth,<br />
dampened with the cleaning solution.<br />
4.5 Sterilization and Cleaning of Reusable<br />
Bladderless Cuffs<br />
Clean cuffs with warm water and a mild detergent. Do not use a detergent containing hand<br />
conditioners, softeners, or fragrances.<br />
NIBP cuffs can be sterilized with gamma sterilization without effecting the repeated<br />
performance of the cuff. Steam sterilization is not recommended. Use of a washing liquid<br />
containing bleach is not recommended because chlorine will chemically break down the<br />
urethane on the inside of the cuff.<br />
Antimicrobial Definition<br />
<strong>Mindray</strong> DS bladderless cuffs are treated with an antimicrobial coating. Antimicrobial<br />
technology effectively controls a broad spectrum of bacteria, fungi, algae and yeasts on a<br />
wide variety of treated substrates.<br />
4 - 2 0070-10-0649-01 <strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions