Decontamination of Monitor User Maintenance<br />

4.3 Decontamination of Monitor<br />

WARNING: Perform the decontamination process with the unit powered<br />

down and power cord removed.<br />

Decontamination of a unit that has come in contact with a biological material can be<br />

performed using LpH SE Germicidal detergent. Apply a small amount of detergent to a<br />

disposable wipe (paper based) and wipe down the outside of the unit. Discard the wipe<br />

appropriately. After waiting 10 minutes, use a clean dry wipe to dry the unit.<br />

CAUTION: During the decontamination process, do not get the LpH SE<br />

Germicidal detergent into any vent openings.<br />

4.4 Care and Cleaning of SpO 2 Sensors<br />

NOTE: Refer to the individual instruction sheets that are packaged<br />

with each sensor.<br />

Daily, check the sensors and cables for signs of damage. Replace as required.<br />

The sensors should be cleaned before and after each patient’s use.<br />

Wipe the patient contact area using a soft cloth with mild soap and water solution or<br />

isopropyl alcohol. Hydrogen peroxide can be used to remove dried blood on all<br />

accessible surfaces.<br />

Let the sensor completely dry before using.<br />

CAUTION: When cleaning SpO 2 sensors, do not use excessive amounts<br />

of liquid. Wipe the sensor surface with a soft cloth,<br />

dampened with the cleaning solution.<br />

4.5 Sterilization and Cleaning of Reusable<br />

Bladderless Cuffs<br />

Clean cuffs with warm water and a mild detergent. Do not use a detergent containing hand<br />

conditioners, softeners, or fragrances.<br />

NIBP cuffs can be sterilized with gamma sterilization without effecting the repeated<br />

performance of the cuff. Steam sterilization is not recommended. Use of a washing liquid<br />

containing bleach is not recommended because chlorine will chemically break down the<br />

urethane on the inside of the cuff.<br />

Antimicrobial Definition<br />

<strong>Mindray</strong> DS bladderless cuffs are treated with an antimicrobial coating. Antimicrobial<br />

technology effectively controls a broad spectrum of bacteria, fungi, algae and yeasts on a<br />

wide variety of treated substrates.<br />

4 - 2 0070-10-0649-01 <strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions