Operation ECG Measurements<br />

Modified Chest Lead (MCL) Monitoring<br />

The recommended lead placement for MCL monitoring is as follows.<br />

Black<br />

LA<br />

FIGURE 3-8 MCL Monitoring with a 3-wire Lead Set (AHA)<br />

RA<br />

LL<br />

Place the RA (white) electrode under the patient’s left clavicle, at the mid-clavicular line<br />

within the rib cage frame.<br />

Place the LA (black) electrode on the right sternal border, at the fourth intercostal space<br />

within the rib cage frame.<br />

Place the LL (red) electrode on the patient’s lower left abdomen within the rib cage frame.<br />

Select ECG Lead I for MCL1 monitoring. Lead I is the direct electrical line between the RA<br />

(white) electrode and the LA (black) electrode.<br />

Select ECG Lead II for MCL6 monitoring. Lead II is the direct electrical line between the RA<br />

(white) electrode and the LL (red) electrode.<br />

<strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 3 - 19<br />