Appendix Newborn NIBP Technique<br />

6.9 Newborn NIBP Technique<br />

Newborn patients present unique obstacles to NIBP measurement. Their vital signs can<br />

change from moment to moment, and their tiny physiological signals are very prone to noise<br />

interference. The following suggestions will help you to obtain the best possible NIBP<br />

measurement.<br />

Try to measure infants when they are calm. A kicking/crying baby may disturb or jiggle<br />

the cuff, causing noise within the system and, as a result, yielding unstable blood pressure<br />

readings. If necessary, hold the cuffed limb steady, but do not impede circulation; do not<br />

hold onto the cuff and do not “pat” the cuffed limb to comfort the child.<br />

Try the calf. Irritable newborns will react to the cuff pressure but may tolerate the calf<br />

better than the arm. Place the cuff just above the ankle.<br />

Use the correct size cuff. <strong>Mindray</strong> DS offers Newborn and Infant size cuffs. When<br />

applying, verify the cuff’s “Index” line falls between the “Range” lines.<br />

Try disposable cuffs. Disposable cuffs are more pliant than the reusable ones. They<br />

generally fit really tiny infants better.<br />

Place the cuff lightly. If the cuff is too snug, it won’t work properly. On infants, you should<br />

be able to easily move the cuff over the limb.<br />

Remember NIBP cannot be taken under all conditions. Even manual methods, employing a<br />

sphygmomanometer and stethoscope, will not work on unstable or active patients.<br />

6.10 How To Get Help<br />

<strong>Mindray</strong> DS maintains a network of service representatives and factory-trained distributors.<br />

Prior to requesting service, perform a complete operational check of the instrument to verify<br />

proper control settings. If operational problems continue to exist, contact the <strong>Mindray</strong> DS<br />

Service Department at (800) 288-2121 or (201) 995-8116 for assistance in determining the<br />

nearest field service location.<br />

Please include the instrument model number, the serial number, and a description of the<br />

problem with all requests for service.<br />

Any questions regarding the warranty should be directed to the nearest <strong>Mindray</strong> DS location.<br />

A list of international offices, along with their phone numbers, is provided at the end of this<br />

manual.<br />

NOTE: Upon request, <strong>Mindray</strong> DS will provide circuit diagrams,<br />

component part lists, descriptions, calibration instructions,<br />

or other information which will assist the user’s<br />

appropriately qualified technical personnel to repair those<br />

parts of the <strong>Mindray</strong> DS equipment which are designated by<br />

<strong>Mindray</strong> DS as repairable.<br />

<strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 6 - 41