Operation Laser Printing 12 Lead ECG (optional - <strong>Passport</strong> 2 only)<br />

3.21 Laser Printing 12 Lead ECG (optional - <strong>Passport</strong> 2<br />

only)<br />

The <strong>Passport</strong> 2 may be connected to a laser printer to print 12 Lead ECG data and<br />

interpretation. Please refer to P/N 0070-00-0441-02 for Laser Printer setup instructions and<br />

refer to section 5.1.9 for the approved printer(s).<br />

3.21.1 Printing 12 Lead to the Laser printer<br />

1. To select the destination for your 12 Lead ECG data, go to the Print Setup menu and<br />

ensure “Select Printer” is set to Laser, Laser/Local or Laser/Remote.<br />

2. To print 12 Lead ECG data to the laser printer, the View 12 Lead ECG Analysis Module<br />

must be installed and enabled within the <strong>Passport</strong> 2.<br />

3. Enter patient's Date of Birth and Gender into the Patient menu to print the 12 Lead<br />

interpretation.<br />

4. Press the VIEW key to display View All ECG mode.<br />

5. Press the STRIP key to print 12 Lead ECG to the destination of choice.<br />

The laser printer will print the patient's demographic data, vital sign data (as for a normal<br />

waveform), and ECG measurement data (Ventricular rate, PR interval, QRS duration, QT/<br />

QTC and P-QRS-T axes) in the upper left portion of the printout. If all conditions are met for<br />

ECG analysis, the 12 Lead interpretation will be printed in the upper right portion of the<br />

printout. Analysis consists of an interpretive statement, a condition statement and a rhythm<br />

statement as specified in the Physician's Guide to Computerized ECG Analysis (<strong>Mindray</strong> DS<br />

P/N 0070-00-0524-01).<br />

Below the digital information and interpretation are 2-½ second samples of each of the 12<br />

ECG vector waveforms displayed in standard 12-Lead format. The waveform samples<br />

represent the 2-½ seconds prior to pressing the STRIP key. Beneath the 12 ECG waveform<br />

samples is a 10-second rhythm strip of ECG Lead II.<br />

The conditions for printing the ECG analysis are:<br />

1. The <strong>Passport</strong> 2 patient size must be set to “Adult.”<br />

2. The patient's Gender and Date of Birth must be entered via the Patient menu.<br />

3. The patient must be at least 18 years old. (The monitor calculates the patient's age from<br />

the Date of Birth entered).<br />

<strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 3 - 77