ECG Measurements Operation<br />

3.4.2 The ECG Menu<br />

FIGURE 3-12 ECG Menu<br />

The ECG Menu provides the following choices: Normal Screen, ECG 1 — ECG 6,<br />

Arrhythmia Menu, Relearn, ST Menu, ECG Sizes Menu, ECG Setup and Resp<br />

Menu.<br />

The Normal Screen selection returns the view to the normal screen.<br />

The ECG 1 — ECG 6 selections define the ECG labels for printing and trends.<br />

The Arrhythmia Menu selection opens the Arrhythmia Menu.<br />

The Relearn selection is only available if the ST or Arrhythmia options are installed and<br />

is used to manually initiate the learning process for ST measurements or Arrhythmia<br />

analysis.<br />

The ST Menu selection opens the ST Menu.<br />

The ECG Sizes Menu selection opens the ECG Sizes Menu.<br />

The ECG Setup selection opens the ECG Setup Menu that is detailed in the following<br />

table.<br />

The Resp Menu selection opens the Resp Menu.<br />

ECG Setup Menu<br />

MENU ITEM SELECTIONS FACTORY DEFAULT/COMMENTS<br />

Previous Menu Returns to the previous menu.<br />

Filter Monitor, Extended, Select to change the filter mode for ECG. Extended or<br />

ST<br />

ST must be used for ST analysis. The filter setting<br />

affects both the display output and the printer output.<br />

Monitor = 0.5 - 40 Hz<br />

Extended, 3 or 5-lead = 0.05 - 100 Hz<br />

Extended, 12-lead = 0.05 - 150 Hz<br />

ST = 0.05 - 40 Hz<br />

3 - 22 0070-10-0649-01 <strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions