Datascope Passport - Mindray
Datascope Passport - Mindray
Datascope Passport - Mindray
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
ECG Measurements Operation<br />
3.4.2 The ECG Menu<br />
FIGURE 3-12 ECG Menu<br />
The ECG Menu provides the following choices: Normal Screen, ECG 1 — ECG 6,<br />
Arrhythmia Menu, Relearn, ST Menu, ECG Sizes Menu, ECG Setup and Resp<br />
Menu.<br />
The Normal Screen selection returns the view to the normal screen.<br />
The ECG 1 — ECG 6 selections define the ECG labels for printing and trends.<br />
The Arrhythmia Menu selection opens the Arrhythmia Menu.<br />
The Relearn selection is only available if the ST or Arrhythmia options are installed and<br />
is used to manually initiate the learning process for ST measurements or Arrhythmia<br />
analysis.<br />
The ST Menu selection opens the ST Menu.<br />
The ECG Sizes Menu selection opens the ECG Sizes Menu.<br />
The ECG Setup selection opens the ECG Setup Menu that is detailed in the following<br />
table.<br />
The Resp Menu selection opens the Resp Menu.<br />
ECG Setup Menu<br />
MENU ITEM SELECTIONS FACTORY DEFAULT/COMMENTS<br />
Previous Menu Returns to the previous menu.<br />
Filter Monitor, Extended, Select to change the filter mode for ECG. Extended or<br />
ST<br />
ST must be used for ST analysis. The filter setting<br />
affects both the display output and the printer output.<br />
Monitor = 0.5 - 40 Hz<br />
Extended, 3 or 5-lead = 0.05 - 100 Hz<br />
Extended, 12-lead = 0.05 - 150 Hz<br />
ST = 0.05 - 40 Hz<br />
3 - 22 0070-10-0649-01 <strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions