Performance Specifications Appendix<br />
6.2.15 Comm-Port<br />
Communication with external devices (i.e. VISA or Patient Net Central Station, Remote<br />
Display, Gas Module and Nurse Call System) is accomplished via the Comm- Port. The<br />
Comm-Port is permanently mounted in a recess on the under side of the monitor.<br />
6.2.15.1 Physical Characteristics<br />
Maximum Size: Has no impact on the size of the<br />
<strong>Passport</strong> 2/<strong>Passport</strong> 2 LT.<br />
Maximum Weight: 0.5 lbs.<br />
6.2.15.2 Communication Characteristics<br />
The Comm Port supports the following connections:<br />
Remote VGA Display - Connection to a medical grade VGA monitor for remote repeater<br />
display P/N 0160-00-0022. The display mimics the main display (15 pin D-subminiature<br />
female connector).<br />
Nurse Call - The Nurse Call function interfaces to hospital nurse call systems. The<br />
connector is a 3-pin DIN connector. The monitor activates a relay, which resides in an<br />
interface cable P/N 0012-00-1277-02. The cable interfaces with “normal open”<br />
signaling systems with a minimum of 1,500VAC isolation.<br />
Serial Communication Channel Supports <strong>Passport</strong> 2 to Visa (with Admit) protocol,<br />
PatientNet protocol, <strong>Mindray</strong> DS Improved ASCII Protocol (DIAP) 1 , Accutorr<br />
Communications Protocols 2 and Gas Module. The interface is proprietary.<br />
Ethernet Communications Port - Supports <strong>Passport</strong> 2 interface to Panorama Central<br />
Station or to laser printer using 100 BASE-TX Ethernet.<br />
1<br />
The <strong>Passport</strong> 2 supports the DIAP (P/N 0070-00-0307) with the following exceptions:<br />
The NIBP elapsed time is set to “—-” when the elapsed time is greater than 999<br />
minutes.<br />
Though not specified in the protocol, the alarm limit values are at the same resolution<br />
as the parameter value. Example, temperature is 10X, therefore, the alarm limit values<br />
are also 10X.<br />
Baud rate is 9600 or 19200.<br />
2The <strong>Passport</strong> 2 supports the Accutorr Communication Protocol (P/N 0070-00-0304) with<br />
the following exceptions:<br />
Baud rate is 9600 or 1200.<br />
The “Show patient data” command fields not supported are:<br />
“f” Status Byte 0<br />
“g” Status Byte 1<br />
“h” Status Byte 2<br />
“I” Status Byte 3<br />
6 - 24 0070-10-0649-01 <strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions