Performance Specifications Appendix<br />

6.2.15 Comm-Port<br />

Communication with external devices (i.e. VISA or Patient Net Central Station, Remote<br />

Display, Gas Module and Nurse Call System) is accomplished via the Comm- Port. The<br />

Comm-Port is permanently mounted in a recess on the under side of the monitor.<br />

6.2.15.1 Physical Characteristics<br />

Maximum Size: Has no impact on the size of the<br />

<strong>Passport</strong> 2/<strong>Passport</strong> 2 LT.<br />

Maximum Weight: 0.5 lbs.<br />

6.2.15.2 Communication Characteristics<br />

The Comm Port supports the following connections:<br />

Remote VGA Display - Connection to a medical grade VGA monitor for remote repeater<br />

display P/N 0160-00-0022. The display mimics the main display (15 pin D-subminiature<br />

female connector).<br />

Nurse Call - The Nurse Call function interfaces to hospital nurse call systems. The<br />

connector is a 3-pin DIN connector. The monitor activates a relay, which resides in an<br />

interface cable P/N 0012-00-1277-02. The cable interfaces with “normal open”<br />

signaling systems with a minimum of 1,500VAC isolation.<br />

Serial Communication Channel Supports <strong>Passport</strong> 2 to Visa (with Admit) protocol,<br />

PatientNet protocol, <strong>Mindray</strong> DS Improved ASCII Protocol (DIAP) 1 , Accutorr<br />

Communications Protocols 2 and Gas Module. The interface is proprietary.<br />

Ethernet Communications Port - Supports <strong>Passport</strong> 2 interface to Panorama Central<br />

Station or to laser printer using 100 BASE-TX Ethernet.<br />

1<br />

The <strong>Passport</strong> 2 supports the DIAP (P/N 0070-00-0307) with the following exceptions:<br />

The NIBP elapsed time is set to “—-” when the elapsed time is greater than 999<br />

minutes.<br />

Though not specified in the protocol, the alarm limit values are at the same resolution<br />

as the parameter value. Example, temperature is 10X, therefore, the alarm limit values<br />

are also 10X.<br />

Baud rate is 9600 or 19200.<br />

2The <strong>Passport</strong> 2 supports the Accutorr Communication Protocol (P/N 0070-00-0304) with<br />

the following exceptions:<br />

Baud rate is 9600 or 1200.<br />

The “Show patient data” command fields not supported are:<br />

“f” Status Byte 0<br />

“g” Status Byte 1<br />

“h” Status Byte 2<br />

“I” Status Byte 3<br />

6 - 24 0070-10-0649-01 <strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions