Arrhythmia Algorithm (Optional <strong>Passport</strong> 2) Operation<br />
3.8 Arrhythmia Algorithm (Optional <strong>Passport</strong> 2)<br />
The <strong>Passport</strong> 2 uses an arrhythmia algorithm to monitor ECG waveform data. The<br />
algorithm creates ECG waveform templates based on a patient’s normal ECG data and uses<br />
them to analyze newly received data. The algorithm verifies that data is free from noise and<br />
artifact, and that it does not deviate from the patient’s normal ECG rhythms.<br />
A normal ECG waveform typically includes consistent spacing between R waves, a sharp<br />
and well defined QRS complex, and an ECG baseline that is free of noise and artifact.<br />
P<br />
ISO Point<br />
Q<br />
FIGURE 3-16 Sample Waveform<br />
Noise and Artifact<br />
R<br />
ST deviation<br />
(Depression or Elevation)<br />
S<br />
J Point<br />
The presence of noise or artifact in an ECG waveform makes the accurate detection and<br />
classification of heart beats difficult. To best optimize performance, all leads should be free<br />
of noise.<br />
Some of the causes of ECG noise include poor skin preparation, improperly attached<br />
electrodes, dried electrode gel, defective leadwires, and patient movement. The algorithm<br />
uses several techniques to differentiate a patient’s QRS complexes from noise sources.<br />
If noise levels are too high, the following will occur until the signal quality is re-established:<br />
Beat detection is suspended<br />
All rhythm calls are suspended<br />
An ECG Noise message is displayed when ECG noise is detected in one or more ECG<br />
leads. If ECG noise continues beyond the configured noise delay, an alarm is triggered,<br />
and ECG rhythm analysis is stopped.<br />
3 - 34 0070-10-0649-01 <strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions<br />
ST Point<br />
ST Segment<br />
T<br />
40 to 80 msec