Arrhythmia Algorithm (Optional <strong>Passport</strong> 2) Operation<br />

3.8 Arrhythmia Algorithm (Optional <strong>Passport</strong> 2)<br />

The <strong>Passport</strong> 2 uses an arrhythmia algorithm to monitor ECG waveform data. The<br />

algorithm creates ECG waveform templates based on a patient’s normal ECG data and uses<br />

them to analyze newly received data. The algorithm verifies that data is free from noise and<br />

artifact, and that it does not deviate from the patient’s normal ECG rhythms.<br />

A normal ECG waveform typically includes consistent spacing between R waves, a sharp<br />

and well defined QRS complex, and an ECG baseline that is free of noise and artifact.<br />

P<br />

ISO Point<br />

Q<br />

FIGURE 3-16 Sample Waveform<br />

Noise and Artifact<br />

R<br />

ST deviation<br />

(Depression or Elevation)<br />

S<br />

J Point<br />

The presence of noise or artifact in an ECG waveform makes the accurate detection and<br />

classification of heart beats difficult. To best optimize performance, all leads should be free<br />

of noise.<br />

Some of the causes of ECG noise include poor skin preparation, improperly attached<br />

electrodes, dried electrode gel, defective leadwires, and patient movement. The algorithm<br />

uses several techniques to differentiate a patient’s QRS complexes from noise sources.<br />

If noise levels are too high, the following will occur until the signal quality is re-established:<br />

Beat detection is suspended<br />

All rhythm calls are suspended<br />

An ECG Noise message is displayed when ECG noise is detected in one or more ECG<br />

leads. If ECG noise continues beyond the configured noise delay, an alarm is triggered,<br />

and ECG rhythm analysis is stopped.<br />

3 - 34 0070-10-0649-01 <strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions<br />

ST Point<br />

ST Segment<br />

T<br />

40 to 80 msec