Operation Non-Invasive Blood Pressure Measurements (NIBP)<br />
3.3.3.2 Suspension of NIBP Measurements<br />
1. Press STOP to suspend an automatically timed measurement sequence or to end a<br />
measurement cycle already in progress (deflate cuff).<br />
2. Press START to take an immediate measurement and resume a suspended timed<br />
measurement sequence.<br />
NOTE: You can press STOP at any time to postpone a scheduled<br />
measurement or to terminate a measurement cycle already<br />
in progress.<br />
CAUTION: Observe caution on all patients (Neonates, Pediatrics, and<br />
Adults) when NIBP is set to the Continuous Mode and the 1<br />
minute interval. When the NIBP “continuous” interval is<br />
chosen, the <strong>Passport</strong> 2/<strong>Passport</strong> 2 LT will continually take<br />
back to back blood pressure readings. As a safety<br />
precaution, a limit is placed on continuous and 1 minute<br />
interval measurements. In continuous mode, after 5<br />
minutes, the NIBP interval will automatically switch to one<br />
measurement taken every 5 minutes. In 1 minute mode,<br />
after 10 minutes the NIBP interval automatically switches to<br />
measurements taken once every 10 minutes. Reports have<br />
been made of nerve injury occurring during use of<br />
automatically cycled blood pressure cuffs. See the<br />
Appendix, “Cautions when Using Automatically Cycled<br />
Blood Pressure Cuffs”.<br />
3.3.4 NIBP Pressure Limit Fail Safe<br />
If the cuff is over-pressurized, the cuff will automatically vent to atmosphere and the NIBP<br />
message window will alternately read “cuff over pressure” and “unable to measure”. Power<br />
the system off and then on again.<br />
3.3.5 Cuff Inflation Time<br />
If the cuff pressure does not attain 20 mmHg within 40 seconds of the start of inflation or if<br />
the target pressure is not reached within another 60 seconds, then the cuff is vented and the<br />
“RETRY” or “UNABLE TO MEASURE” message will display in the NIBP message window.<br />
3.3.6 START and STOP Functions<br />
The START and STOP functions have the following effects on the timed measurement<br />
sequence (Interval or Timer Mode).<br />
INTERVAL is set and you press START:<br />
An unscheduled measurement is made. Taking this unscheduled measurement does not<br />
affect the timing of the interval cycle, therefore, the scheduled measurements will be taken<br />
as if there were no interruptions. Only one measurement is taken for each measurement<br />
cycle - therefore, if the unscheduled measurement coincides with the scheduled<br />
measurement, it counts as the scheduled measurement.<br />
INTERVAL is set and you press STOP during the measurement:<br />
The cuff deflates and interval measurements are suspended.<br />
INTERVAL is set and you change the interval:<br />
The measurement cycle is reset with the new interval. A measurement will be taken after<br />
you press the START key.<br />
<strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 3 - 11