Operation Non-Invasive Blood Pressure Measurements (NIBP)<br />

3.3.3.2 Suspension of NIBP Measurements<br />

1. Press STOP to suspend an automatically timed measurement sequence or to end a<br />

measurement cycle already in progress (deflate cuff).<br />

2. Press START to take an immediate measurement and resume a suspended timed<br />

measurement sequence.<br />

NOTE: You can press STOP at any time to postpone a scheduled<br />

measurement or to terminate a measurement cycle already<br />

in progress.<br />

CAUTION: Observe caution on all patients (Neonates, Pediatrics, and<br />

Adults) when NIBP is set to the Continuous Mode and the 1<br />

minute interval. When the NIBP “continuous” interval is<br />

chosen, the <strong>Passport</strong> 2/<strong>Passport</strong> 2 LT will continually take<br />

back to back blood pressure readings. As a safety<br />

precaution, a limit is placed on continuous and 1 minute<br />

interval measurements. In continuous mode, after 5<br />

minutes, the NIBP interval will automatically switch to one<br />

measurement taken every 5 minutes. In 1 minute mode,<br />

after 10 minutes the NIBP interval automatically switches to<br />

measurements taken once every 10 minutes. Reports have<br />

been made of nerve injury occurring during use of<br />

automatically cycled blood pressure cuffs. See the<br />

Appendix, “Cautions when Using Automatically Cycled<br />

Blood Pressure Cuffs”.<br />

3.3.4 NIBP Pressure Limit Fail Safe<br />

If the cuff is over-pressurized, the cuff will automatically vent to atmosphere and the NIBP<br />

message window will alternately read “cuff over pressure” and “unable to measure”. Power<br />

the system off and then on again.<br />

3.3.5 Cuff Inflation Time<br />

If the cuff pressure does not attain 20 mmHg within 40 seconds of the start of inflation or if<br />

the target pressure is not reached within another 60 seconds, then the cuff is vented and the<br />

“RETRY” or “UNABLE TO MEASURE” message will display in the NIBP message window.<br />

3.3.6 START and STOP Functions<br />

The START and STOP functions have the following effects on the timed measurement<br />

sequence (Interval or Timer Mode).<br />

INTERVAL is set and you press START:<br />

An unscheduled measurement is made. Taking this unscheduled measurement does not<br />

affect the timing of the interval cycle, therefore, the scheduled measurements will be taken<br />

as if there were no interruptions. Only one measurement is taken for each measurement<br />

cycle - therefore, if the unscheduled measurement coincides with the scheduled<br />

measurement, it counts as the scheduled measurement.<br />

INTERVAL is set and you press STOP during the measurement:<br />

The cuff deflates and interval measurements are suspended.<br />

INTERVAL is set and you change the interval:<br />

The measurement cycle is reset with the new interval. A measurement will be taken after<br />

you press the START key.<br />

<strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 3 - 11