Appendix Performance Specifications<br />
6.2.10.2 CO 2 Respiration Performance Requirements<br />
6.2.11 SpO 2<br />
MediCO 2 Microstream ® (Only in monitors with serial numbers<br />
below TS10000.)<br />
Respiration Rate Range: 0 - 150 RPM.<br />
Respiration Rate Accuracy: ± 1 RPM from 0 to 40 RPM<br />
MiniMediCO 2 Microstream ® (Only in monitors with serial number<br />
TS10000 and higher.)<br />
Respiration Rate Range: 0 - 150 RPM.<br />
SpO2 is a non-invasive measurement of the functional oxygen saturation.<br />
The SPO2 function is in accordance with the requirements of EN 865.<br />
6.2.11.1 Masimo ® SpO 2 Performance Requirements<br />
± 2 RPM from 41 to 70 RPM<br />
± 3 RPM from 71 to 100 RPM<br />
± 5 RPM from 101 to 150 RPM<br />
Respiration Rate Accuracy: ± 1 RPM from 0 to 70 RPM<br />
Sensor Compatibility:<br />
Masimo: LNOP ® Series.<br />
± 2 RPM from 71 to 120 RPM<br />
± 3 RPM from 121 to 150 RPM<br />
Connector Type: 14-lead “Mini D-Ribbon” Receptacle (Masimo<br />
PC-12).<br />
Measurement Technique:<br />
Masimo Sensors: Masimo proprietary algorithm (SET).<br />
SpO 2 Accuracy Saturation during No Motion Conditions 1 :<br />
Adults / Pediatrics: 70% to 100% ± 2 5 , 0 to 69% unspecified.<br />
Neonates: 70% to 100% ± 3, 0% to 69% unspecified.<br />
SpO 2 Accuracy Saturation during Motion Conditions 6 :<br />
Adults / Pediatrics 2 : 70% to 100% ± 3, 0 to 69% unspecified.<br />
<strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 6 - 17