Monitor Problem Solving Operation<br />
PROBLEM REASON ACTION<br />
Display Appears to be<br />
Off<br />
Disabled Alarm Tone,<br />
QRS Tone, or Other<br />
Function<br />
ECG Base Line With No<br />
Waveform<br />
Mains power switch may not be<br />
on<br />
Unit may not be plugged into an<br />
AC outlet<br />
If used as a portable, battery<br />
pack may be drained<br />
Check mains power switch on side<br />
panel<br />
Check power cord (Is it plugged<br />
in?)<br />
If battery pack is drained, plug<br />
into an AC outlet to recharge the<br />
battery. A period of 5 hours is<br />
required for a full charge of<br />
lithium-ion batteries. A period of<br />
16 hours is required for a full<br />
charge of sealed lead acid<br />
batteries.<br />
MUTE key pressed Check for alarm mute symbol and<br />
message<br />
Beep volume low Increase beep volume<br />
Gain control not set high enough.<br />
Set through SIZE key<br />
Leadwires and patient cable not<br />
fully inserted into proper<br />
receptacle<br />
Readjust as required<br />
Check insertion<br />
Cable or leadwires damaged Check with lead continuity tester<br />
Base Line Wander Patient moving excessively Secure leadwires and cable to<br />
patient<br />
Patient's respiration Reposition electrodes<br />
Electrodes dry Re-prep skin and apply fresh moist<br />
electrodes<br />
Static build up around patient Check with hospital engineer<br />
Damped Invasive<br />
Air bubbles in tubing Eliminate air from tubing<br />
Waveform<br />
Kinked catheter Slightly alter position of catheter<br />
Catheter against wall of blood<br />
vessel<br />
Check for leaks at connector<br />
Recorder Report Appears<br />
Totally Blank<br />
Resp. Waveform Too<br />
Large<br />
Resp. Waveform Too<br />
Small<br />
Blood in tubing Pump pressure bag up to 300<br />
mmHg<br />
Thermal paper may be installed<br />
incorrectly (up-side down)<br />
Remove paper and re-install with<br />
paper feeding off of the spool<br />
from the bottom<br />
Scales set inappropriately Change lead selection<br />
Patient breathing shallow or<br />
turned on side<br />
Scale set inappropriately<br />
False Apnea Alarm Apnea delay may be improperly<br />
set<br />
Patient may be having frequent<br />
episodes of CVA<br />
Scale size may be too low<br />
Change lead selection<br />
Choose another apnea delay<br />
Reposition electrodes to better<br />
detect respirations<br />
3 - 86 0070-10-0649-01 <strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions