Monitor Problem Solving Operation<br />

PROBLEM REASON ACTION<br />

Display Appears to be<br />

Off<br />

Disabled Alarm Tone,<br />

QRS Tone, or Other<br />

Function<br />

ECG Base Line With No<br />

Waveform<br />

Mains power switch may not be<br />

on<br />

Unit may not be plugged into an<br />

AC outlet<br />

If used as a portable, battery<br />

pack may be drained<br />

Check mains power switch on side<br />

panel<br />

Check power cord (Is it plugged<br />

in?)<br />

If battery pack is drained, plug<br />

into an AC outlet to recharge the<br />

battery. A period of 5 hours is<br />

required for a full charge of<br />

lithium-ion batteries. A period of<br />

16 hours is required for a full<br />

charge of sealed lead acid<br />

batteries.<br />

MUTE key pressed Check for alarm mute symbol and<br />

message<br />

Beep volume low Increase beep volume<br />

Gain control not set high enough.<br />

Set through SIZE key<br />

Leadwires and patient cable not<br />

fully inserted into proper<br />

receptacle<br />

Readjust as required<br />

Check insertion<br />

Cable or leadwires damaged Check with lead continuity tester<br />

Base Line Wander Patient moving excessively Secure leadwires and cable to<br />

patient<br />

Patient's respiration Reposition electrodes<br />

Electrodes dry Re-prep skin and apply fresh moist<br />

electrodes<br />

Static build up around patient Check with hospital engineer<br />

Damped Invasive<br />

Air bubbles in tubing Eliminate air from tubing<br />

Waveform<br />

Kinked catheter Slightly alter position of catheter<br />

Catheter against wall of blood<br />

vessel<br />

Check for leaks at connector<br />

Recorder Report Appears<br />

Totally Blank<br />

Resp. Waveform Too<br />

Large<br />

Resp. Waveform Too<br />

Small<br />

Blood in tubing Pump pressure bag up to 300<br />

mmHg<br />

Thermal paper may be installed<br />

incorrectly (up-side down)<br />

Remove paper and re-install with<br />

paper feeding off of the spool<br />

from the bottom<br />

Scales set inappropriately Change lead selection<br />

Patient breathing shallow or<br />

turned on side<br />

Scale set inappropriately<br />

False Apnea Alarm Apnea delay may be improperly<br />

set<br />

Patient may be having frequent<br />

episodes of CVA<br />

Scale size may be too low<br />

Change lead selection<br />

Choose another apnea delay<br />

Reposition electrodes to better<br />

detect respirations<br />

3 - 86 0070-10-0649-01 <strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions