Alarms Operation<br />
3.19.2 Alarm Limits<br />
NOTE: Alarm limits can be set using the “Auto-Set” function. Alarm<br />
limits can be saved in the Monitor Setup Menu. The “Save<br />
Current” option must be selected in order to save the current<br />
parameters.<br />
All of the alarm limits have an “OFF” position with the exception of low SpO2, high inspired<br />
N2O and low inspired O2 . A separate table of alarm limit settings is maintained for each<br />
patient size. When the patient size is changed, the appropriate selections are automatically<br />
used. See table below for alarm ranges.<br />
Alarm Parameters<br />
HIGH LO<br />
PARA<br />
METERS ADULT PED NEONATE ADULT PED NEONATE<br />
Heart Rate<br />
(bpm)<br />
Off,<br />
60-250<br />
Off,<br />
100-300<br />
Off,<br />
100-350<br />
Off,<br />
30-120<br />
Off,<br />
30-150<br />
PVC/ min Off, 1-30<br />
Run** 4 / 4-8<br />
ST Single<br />
Lead (mm)<br />
Off, +0.5 - +10.0 (Elevation) Off, -0.5 - -10.0 (Depression)<br />
ST Dual<br />
Lead (mm)<br />
NIBP Sys<br />
(mmHg)<br />
NIBP Mean<br />
(mmHg)<br />
NIBP Dia<br />
(mmHg)<br />
IBP1/IBP2<br />
Sys (mmHg)<br />
IBP1/IBP2<br />
Mean<br />
(mmHg)<br />
IBP1/IBP2<br />
Dia (mmHg)<br />
Off, +0.5 - +10.0 (Elevation) Off, -0.5 - -10.0 (Depression)<br />
Off,<br />
70-240<br />
Off,<br />
60-200<br />
Off,<br />
40-130<br />
Off,<br />
5-300<br />
Off,<br />
5 - 150<br />
Off,<br />
0-140<br />
SpO 2 (%) Off,<br />
80-100<br />
Off,<br />
40-180<br />
Off,<br />
50-180<br />
Off,<br />
50-100<br />
Off,<br />
5-240<br />
Off,<br />
5 -100<br />
Off,<br />
0-100<br />
Off,<br />
80-100<br />
Off,<br />
40-180<br />
Off,<br />
40-160<br />
Off,<br />
50-100<br />
Off,<br />
5-180<br />
Off,<br />
5 - 100<br />
Off,<br />
50-150<br />
Off,<br />
40-140<br />
Off,<br />
30-120<br />
Off,<br />
0-150<br />
Off,<br />
2-100<br />
Off,<br />
15-130<br />
Off,<br />
10-100<br />
Off,<br />
10-50<br />
Off,<br />
0-130<br />
Off,<br />
2 - 50<br />
Off,<br />
30-200<br />
Off,<br />
15-130<br />
Off,<br />
10-70<br />
Off,<br />
10-50<br />
Off,<br />
0-130<br />
3 - 66 0070-10-0649-01 <strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions<br />
Off,<br />
0-70<br />
Off,<br />
80-100<br />
Off,<br />
0-120<br />
Off,<br />
0-100<br />
Off,<br />
2-50<br />
Off,<br />
0-50<br />
50-99 50-99 50-99<br />
T1 (°F) Off, 95-110 Off, 80-100<br />
T1 (°C) Off, 35-43 Off, 26-38<br />
Resp Rate Off, Off, Off,<br />
Off, Off,<br />
(rpm)<br />
10-100 15-150 30-200 5-30 5-40<br />
ET CO2 (mmHg)<br />
Off, 20-80 Off, 5-50<br />
ET CO2 (%) Off, 2.0-10.0 Off, 1.0-6.0<br />
ET CO2 (kPa) Off, 2.0-10.0 Off, 1.0-6.0<br />
Insp CO2 (mmHg)<br />
Off, 5-30<br />
** <strong>Passport</strong> 2 only with 3-lead or 5-lead.<br />
Off,<br />
5-50