Alarms Operation<br />

3.19.2 Alarm Limits<br />

NOTE: Alarm limits can be set using the “Auto-Set” function. Alarm<br />

limits can be saved in the Monitor Setup Menu. The “Save<br />

Current” option must be selected in order to save the current<br />

parameters.<br />

All of the alarm limits have an “OFF” position with the exception of low SpO2, high inspired<br />

N2O and low inspired O2 . A separate table of alarm limit settings is maintained for each<br />

patient size. When the patient size is changed, the appropriate selections are automatically<br />

used. See table below for alarm ranges.<br />

Alarm Parameters<br />

HIGH LO<br />

PARA<br />

METERS ADULT PED NEONATE ADULT PED NEONATE<br />

Heart Rate<br />

(bpm)<br />

Off,<br />

60-250<br />

Off,<br />

100-300<br />

Off,<br />

100-350<br />

Off,<br />

30-120<br />

Off,<br />

30-150<br />

PVC/ min Off, 1-30<br />

Run** 4 / 4-8<br />

ST Single<br />

Lead (mm)<br />

Off, +0.5 - +10.0 (Elevation) Off, -0.5 - -10.0 (Depression)<br />

ST Dual<br />

Lead (mm)<br />

NIBP Sys<br />

(mmHg)<br />

NIBP Mean<br />

(mmHg)<br />

NIBP Dia<br />

(mmHg)<br />

IBP1/IBP2<br />

Sys (mmHg)<br />

IBP1/IBP2<br />

Mean<br />

(mmHg)<br />

IBP1/IBP2<br />

Dia (mmHg)<br />

Off, +0.5 - +10.0 (Elevation) Off, -0.5 - -10.0 (Depression)<br />

Off,<br />

70-240<br />

Off,<br />

60-200<br />

Off,<br />

40-130<br />

Off,<br />

5-300<br />

Off,<br />

5 - 150<br />

Off,<br />

0-140<br />

SpO 2 (%) Off,<br />

80-100<br />

Off,<br />

40-180<br />

Off,<br />

50-180<br />

Off,<br />

50-100<br />

Off,<br />

5-240<br />

Off,<br />

5 -100<br />

Off,<br />

0-100<br />

Off,<br />

80-100<br />

Off,<br />

40-180<br />

Off,<br />

40-160<br />

Off,<br />

50-100<br />

Off,<br />

5-180<br />

Off,<br />

5 - 100<br />

Off,<br />

50-150<br />

Off,<br />

40-140<br />

Off,<br />

30-120<br />

Off,<br />

0-150<br />

Off,<br />

2-100<br />

Off,<br />

15-130<br />

Off,<br />

10-100<br />

Off,<br />

10-50<br />

Off,<br />

0-130<br />

Off,<br />

2 - 50<br />

Off,<br />

30-200<br />

Off,<br />

15-130<br />

Off,<br />

10-70<br />

Off,<br />

10-50<br />

Off,<br />

0-130<br />

3 - 66 0070-10-0649-01 <strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions<br />

Off,<br />

0-70<br />

Off,<br />

80-100<br />

Off,<br />

0-120<br />

Off,<br />

0-100<br />

Off,<br />

2-50<br />

Off,<br />

0-50<br />

50-99 50-99 50-99<br />

T1 (°F) Off, 95-110 Off, 80-100<br />

T1 (°C) Off, 35-43 Off, 26-38<br />

Resp Rate Off, Off, Off,<br />

Off, Off,<br />

(rpm)<br />

10-100 15-150 30-200 5-30 5-40<br />

ET CO2 (mmHg)<br />

Off, 20-80 Off, 5-50<br />

ET CO2 (%) Off, 2.0-10.0 Off, 1.0-6.0<br />

ET CO2 (kPa) Off, 2.0-10.0 Off, 1.0-6.0<br />

Insp CO2 (mmHg)<br />

Off, 5-30<br />

** <strong>Passport</strong> 2 only with 3-lead or 5-lead.<br />

Off,<br />

5-50