Alarms Operation<br />

MUTE - This key is a single action button, which when pressed, silences the alarm<br />

whose parameters have been violated for a programmed length of time (Default 2 minutes),<br />

or until the alarm condition is no longer present, which ever is shorter. Any new alarms that<br />

occur during the silenced period will disable the silence and the alarm will sound the tone.<br />

An Alarm Mute icon, resembling a crossed speaker, is displayed next to each muted<br />

parameter. A message and digital timer counts down in the upper message area. Pressing<br />

MUTE again does not re-enable audio alarms<br />

MUTE ALL - This is a single action button, which when pressed, suspends alarms on<br />

all parameters for a programmed period of time. (Default is 2 minutes). An Alarm Mute icon<br />

(resembling a crossed speaker) is displayed next to each parameter. A message and timer<br />

appear in the upper message area showing the time remaining. Pressing MUTE ALL at any<br />

time re-enables audio alarm tones. If MUTE ALL was permanently selected a message is<br />

posted stating All alarms muted permanently.<br />

The time period for MUTE and MUTE ALL is adjustable via the ALARMS SETUP Menu.<br />

3 - 70 0070-10-0649-01 <strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions