Performance Specifications Appendix<br />

Sensor Compatibility:<br />

Neonates 3 : 70% to 100% ± 3, 0% to 69% unspecified.<br />

SpO2 Response Time: 18 seconds to 95% of final step change of %<br />

SpO2 value from 60 to 95% at 75bpm.<br />

NOTE: This time was measured with post average time set at 8<br />

seconds.<br />

Low Perfusion Performance 4 : >0.02% Pulse Amplitude and % Transmission<br />

>5%.<br />

Saturation (%SpO2 ) ± 2 digits, Pulse ± 3<br />

digits.<br />

Wavelengths Emitted: 660 nm and 905 nm.<br />

Maximum Emitted Energy: 30 mW at 50 mA pulsed.<br />

Interfering Substances: Carboxyhemoglobin may erroneously<br />

increase readings. The level of increase is<br />

approximately equal to the amount of<br />

carboxyhemoglobin present. Dyes, or any<br />

substance containing dyes, that change usual<br />

arterial pigmentation may cause erroneous<br />

readings.<br />

6 - 18 0070-10-0649-01 <strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions