Appendix Performance Specifications<br />
Optionally, non-lethal or all arrhythmia alarms are capable of being disabled.<br />
In 3 Lead and 5 Lead ECG modes: PVCs per minute are invalidated and the PVC counter<br />
is reset to 0 during periods of Ventricular Rhythm, Ventricular Tachycardia, Ventricular<br />
Fibrillation or Asystole.<br />
In 12 Lead ECG mode: PVCs per minute are invalidated and the PVC counter is reset to 0<br />
during periods of Ventricular Rhythm, Ventricular Tachycardia, and Ventricular Fibrillation.<br />
During periods of Asystole, the PVC counter is reset to 0.<br />
6.2.5.1 12-Lead ECG Interpretation<br />
The <strong>Passport</strong> 2, equipped with optional 12-Lead ECG, has the ability to make<br />
Interpretation, Rhythm and Condition statements in accordance with the Physician’s Guide to<br />
Computerized ECG Analysis (<strong>Mindray</strong> DS P/N 0070-00-0524-01 English, 0070-00-0524-<br />
50 all other languages).<br />
Interpretive statements are disabled in neonatal and pediatric modes.<br />
Interpretive statements are disabled in adult mode for ages less than 18 years old.<br />
6.2.6 NIBP Sub-System Performance Characteristics<br />
This function calculates the patient blood pressure non-invasively, using a blood pressure cuff.<br />
There are three pressure readings available: systolic, diastolic, and mean pressures.<br />
Systolic Pressure Measurements<br />
Accuracy 1,2 : Mean error is less than ± 5 mmHg<br />
Diastolic Pressure Measurements<br />
Standard Deviation shall be less than ± 8<br />
mmHg.<br />
Range: 55 to 235 mmHg in Adult mode<br />
55 to 160 mmHg in Pediatric mode.<br />
45 to 120 mmHg in Neonatal mode.<br />
Accuracy 1,2 : Mean error is less than ± 5 mmHg,<br />
Standard deviation is less than ± 8 mmHg.<br />
Range: 30 to 200 mmHg in Adult mode<br />
30 to 150 mmHg in Pediatric mode.<br />
20 to 100 mmHg in Neonatal mode.<br />
1Adult and Pediatric - Blood pressure measurements determined with this device are<br />
equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation<br />
method, within the limits prescribed by the American National Standard, Electronic or<br />
automated sphygmomanometers.<br />
<strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 6 - 11