SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>SonTek</strong><br />
2<br />
• Click Realtime to start the real-time data collection software.<br />
• Select a short averaging interval (10 seconds)<br />
• Press OK to configure the <strong>Argonaut</strong>.<br />
• Press the green play icon (!) to start data collection.<br />
• Press the red record icon (") to start recording to a file.<br />
• Collect a few minutes worth of data, generating real currents if possible.<br />
4) Quick Start: Post Processing with Windows Software<br />
• Start View<strong>Argonaut</strong> using Start | Programs | <strong>SonTek</strong> Software | View<strong>Argonaut</strong>.<br />
• Click PostProcessing to start the post processing software.<br />
• Select File | Open.<br />
• Select the file you recorded with the real-time software and click OK.<br />
• Click OK when file statistics are shown to finish loading the data.<br />
• Use the Zoom tool.<br />
• Double click the axis to change the plot scale.<br />
• Use the slide bar on the top to view tabular data for individual samples.<br />
• Export data in ASCII format using File | Export.<br />
5) Quick Start: System Diagnostics with Windows Software<br />
• Mount the <strong>Argonaut</strong> in a small tank of water with a little seeding material, ideally with the<br />
instrument close to a boundary (20-30 cm / 8-12 in).<br />
• Start View<strong>Argonaut</strong> using Start | Programs | <strong>SonTek</strong> Software | View<strong>Argonaut</strong>.<br />
• Click ArgCheck to start the system diagnostic software.<br />
• Refer to the <strong>Argonaut</strong> Software <strong>Manual</strong> for interpreting the software output.<br />
6) Direct Command Interface with Windows Software<br />
• The direct command interface allows you to “talk” directly to the <strong>Argonaut</strong>.<br />
• Start SonUtils using Start | Programs | <strong>SonTek</strong> Software | SonUtils.<br />
• Select SonTermW to start the terminal emulator software.<br />
• Click the BREAK icon (or press Alt+B) to wake-up the instrument. You should see a wakeup<br />
response from the instrument.<br />
• Use the Show icons to view all parameter settings.<br />
• Type help and press for assistance with the direct command interface.<br />
• Refer to the <strong>Argonaut</strong> Operation <strong>Manual</strong> for more details.<br />
Special Notice when Turning Computers On or Off<br />
When many computers and data loggers are turned on or off, they send a signal out the serial port<br />
that can be interpreted as a BREAK. If the <strong>Argonaut</strong> is connected to the serial port, this can<br />
interrupt data collection or awaken the <strong>Argonaut</strong> from its sleep mode. This may cause the loss of<br />
data or the draining of batteries. Always disconnect the <strong>Argonaut</strong> from the computer or data<br />
logger before turning the computer/data logger on or off.<br />
Getting Started with the <strong>Argonaut</strong> MD, SL, or <strong>XR</strong> (May 1, 2001)