4.3. CompCalW Compass Calibration<br />

CompCalW is used to calibrate the internal compass/tilt sensor for the effects of ambient<br />

magnetic fields.<br />

38<br />

• CompCalW establishes communication and starts a display of compass data.<br />

• CompCalW gives calibration instructions and displays compass data.<br />

• At the end of calibration, CompCalW reports a calibration score. See the <strong>Argonaut</strong><br />

Operation <strong>Manual</strong> for information about compass operation and calibration.<br />

• CompCalW records all compass calibration data and results to a log file.<br />

Basic<br />

Communication<br />

Parameters<br />

Compass<br />

Data Display<br />

Main Program<br />

Functions<br />

Program Status Display<br />

RS485<br />

Address Setting<br />

Figure 12 – CompCalW (Compass Calibration) Main Window<br />

<strong>Argonaut</strong> Software <strong>Manual</strong> (March 1, 2001)