To set SDI-12 SIDEKICK format use (you may also use SONTEK in firmware v7.9 or later):<br />

>sdi12format SIDEKICK <br />

If you need the data to be stored on the <strong>Argonaut</strong>’s internal recorder (in addition to being logged<br />

onto an SDI-12 data-logger), turn the recorder ON using:<br />

>recorder on <br />

After setup is complete, save your settings using:<br />

>savesetup <br />

To put the <strong>Argonaut</strong> into SDI-12 data collection mode use:<br />

>SDI12 ON <br />

The <strong>Argonaut</strong> sends this reply to indicate it is now in SDI-12 mode (see Important Notes #1):<br />

OK<br />

Checking Setup Parameters...<br />

917504 free bytes left in recorder.<br />

Free space is sufficient for 2.31 days of operation.<br />

Recorder mode is NORMAL.<br />

Switching to SDI-12 mode<br />

SDI-12 address is: 1<br />

You should verify that the <strong>Argonaut</strong> can now accept SDI-12 commands by sending the following<br />

SDI-12 “Send Identification” command and receiving the appropriate response (§A2.5.7).<br />

?I! (You may be required to enter a also; see Important Notes #2 first.)<br />

112 <strong>SonTek</strong> SL 79 E272<br />

You can now enter additional SDI-12 commands. When done, make sure you disconnect the RS-<br />

SDI switching plug (Figure A2.6) to complete the process of placing <strong>Argonaut</strong> in SDI-12 mode.<br />

Important Notes:<br />

1. If the recorder is enabled and already full, the <strong>Argonaut</strong> responds with an error message and<br />

will not enter SDI-12 mode. To proceed, the recorder has to be either disabled or formatted.<br />

2. Depending on the terminal emulator you are using (SonTerm [for DOS] or SonTermW [for<br />

Windows]), you may be required to enter a to complete the command string.<br />

• SonTerm (for DOS) – Do not enter the terminating . The system recognizes the “!”<br />

character as the command terminator. If you inadvertently send a , you can return<br />

the system to a known state by sending a BREAK, or any other valid SDI-12 command not<br />

followed by a .<br />

• SonTermW (for Windows) – You do enter the terminating , assuming you have set<br />

the File|Communications Mode to SDI-12. If this item is set to RS232/422/485, the system<br />

will act the same as SonTerm for DOS (as described above).<br />

3. If a BREAK is issued to an <strong>Argonaut</strong> that is in SDI-12 mode, the system will not stop its<br />

operation and will send the following message.<br />

100<br />

112 <strong>SonTek</strong> SL 79 E272<br />

System is in SDI-12 mode. To switch to RS232 type ?EXIT!<br />

4. While in SDI-12 mode, the serial communications do not support full duplex. Therefore, the<br />

characters you enter are not echoed back to the screen.<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)