3.17. Output Data Format (Profile Data)<br />

Systems manufactured with the optional Profiling Mode feature will output profile data for each<br />

cell. The format of the data is shown in this section. Before profile data can be displayed, you<br />

must set the ProfilingMode (PM) command to YES. You must also make entries for number of<br />

cells (NC), cell size (CS), and blanking distance (BD). See §3.8 for command descriptions.<br />

The following example and table show the format of <strong>Argonaut</strong> profile data.<br />

38<br />

1 -251 -39 23 0 0 0 31 59 72<br />

2 129 155 -20 0 0 0 31 59 71<br />

3 0 2186 56 0 0 0 0 0 0<br />

4 -267 88 134 0 0 0 31 59 72<br />

<strong>Argonaut</strong> Profile Data Format<br />

Column Contents<br />

1 Cell number<br />

2 Velocity data (V1 or Vx or Veast) in cm/s<br />

3 Velocity data (V2 or Vy or Vnorth) in cm/s<br />

4 Velocity data (V3 or Vz or Vup) in cm/s [3-beam systems only]<br />

5 Standard deviation data (SD1 or SDx or SDeast in cm/s<br />

6 Standard deviation data (SD2 or SDy or SDnorth in cm/s<br />

7 Standard deviation data (SD3 or SDz or SDup in cm/s [3-beam systems only]<br />

8 Amplitude (signal strength), Beam 1 in counts (1 count = 0.43 dB)<br />

9 Amplitude (signal strength), Beam 2 in counts<br />

10 Amplitude (signal strength), Beam 3 in counts [3-beam systems only]<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)