3.5.1. Real-time User Setup Window<br />

After starting, the real-time software shows a blank data collection screen (Figure 7).<br />

<strong>SonTek</strong>/<strong>YSI</strong><br />

• Establish communication with the system by clicking the Arg icon in the upper-left corner<br />

of the screen or use the System Communication menu.<br />

• The software will establish communication (using the communication parameters shown<br />

on the View<strong>Argonaut</strong> main menu, §3.1) and download the system configuration.<br />

• The software displays the real-time setup window and lets you modify operating<br />

parameters as needed.<br />

Figure 6 – View<strong>Argonaut</strong> Real-Time Data Collection Setup Window<br />

Figure 6 shows the real-time setup window. Each parameter is discussed in detail below.<br />

Temperature Mode:<br />

• Temperature Mode determines the source of temperature data used for sound speed<br />

calculations. Sound speed is used for Doppler velocity calculations (see <strong>Argonaut</strong><br />

Principles of Operation).<br />

• The default setting of MEASURED uses the <strong>Argonaut</strong> temperature sensor, and is<br />

recommended for most applications.<br />

• The USER setting uses the user-supplied value shown in the menu.<br />

<strong>Argonaut</strong> Software <strong>Manual</strong> (March 1, 2001) 19