01.06.2013 Views

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>SonTek</strong>/<strong>YSI</strong><br />

To configure the Converter for operation with the <strong>Argonaut</strong>, use the following command<br />

sequence. A description of the Converter’s setup word is shown in the table below.<br />

116<br />

1. Read the existing setup word.<br />

$1RS<br />

*31070180<br />

2. Write-enable the Converter EEPROM.<br />

$1WE<br />

*<br />

3. Send a new setup word.<br />

$1SU31820180<br />

*<br />

4. Read the new setup word to confirm it is correct.<br />

$1RS<br />

*31820180<br />

5. Write-enable the Converter EEPROM, and then store the changes.<br />

$1WE<br />

*<br />

$1RR<br />

*<br />

Converter’s Setup Word Description<br />

Byte Purpose Setting Description<br />

1 Address 31 Address 1<br />

32 Address 2<br />

33 Address 3<br />

34 Address 4<br />

35 Address 5<br />

36 Address 6<br />

37 Address 7<br />

38 Address 8<br />

2 COM Setting 82 9600 baud, no parity, with linefeed (default)<br />

02 9600 baud, no parity, no linefeed (optional)<br />

3 Advanced COM Setting 01 2-byte time delay, continuous disabled, limits<br />

disabled, no echo (default)<br />

4 Display Settings/Mode Select 80 6-digit display, manual mode disabled, up/down<br />

mode (default)<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)