3.8. Setup Commands<br />
Setup commands affect the primary data collection parameters of the <strong>Argonaut</strong>.<br />
18<br />
• Temperature ● Cell begin<br />
• Salinity ● Cell end<br />
• Averaging interval ● Profiling mode<br />
• Sample interval ● Number of cells<br />
• Velocity Range ● Cell size<br />
• Coordinate system ● Blanking distance<br />
• Data format ● Dynamic boundary adjustment<br />
Note that the timing commands AvgInterval and SampleInterval are also listed under<br />
deployment commands (§3.9).<br />
Each command is shown with its full name, short name (if one exists), optional parameters (in<br />
brackets), parameter range (if appropriate), and detailed information.<br />
Temp [d.d]<br />
• Default parameter: 20.0 (°C)<br />
• Parameter range: -5.0 to 60.0 (°C)<br />
• Without parameter: returns its current setting.<br />
• When used with a valid input parameter, the user-specified temperature is set to this value.<br />
• This temperature value is used to calculate sound speed if TempMode is set to USER.<br />
• See the <strong>Argonaut</strong> Principles of Operation for the effect of sound speed on velocity data.<br />
Sal [d.d]<br />
• Default parameter: 34.5 (ppt)<br />
• Parameter range: 0.0 to 60.0 (ppt)<br />
• Without parameter: returns its current setting.<br />
• When used with a valid input parameter, the salinity is set to this value.<br />
• This salinity value is used to calculate sound speed.<br />
• See the <strong>Argonaut</strong> Principles of Operation for the effect of sound speed on velocity data.<br />
AvgInterval or AI [d]<br />
• Default parameter: 120 (s)<br />
• Parameter range: 10 to 3600 (s; maximum is equivalent to 60 minutes)<br />
• Without parameter: returns its current setting (in seconds).<br />
• When used with a valid integer parameter, the averaging interval is set to this value.<br />
SampleInterval or SI [d]<br />
• Default parameter: 1200 (s; 20 minutes)<br />
• Parameter range: 10 to 43200 (s; maximum is equivalent to 12 hours)<br />
• Without parameter: returns its current setting (in seconds).<br />
• When used with a valid integer parameter, the sample interval is set to this value.<br />
• Sample interval is the time between the start of successive samples (§1.2 and §1.3).<br />
• If AvgInterval > SampleInterval, then AvgInterval takes precedence.<br />
<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)