3.8. Setup Commands<br />

Setup commands affect the primary data collection parameters of the <strong>Argonaut</strong>.<br />

18<br />

• Temperature ● Cell begin<br />

• Salinity ● Cell end<br />

• Averaging interval ● Profiling mode<br />

• Sample interval ● Number of cells<br />

• Velocity Range ● Cell size<br />

• Coordinate system ● Blanking distance<br />

• Data format ● Dynamic boundary adjustment<br />

Note that the timing commands AvgInterval and SampleInterval are also listed under<br />

deployment commands (§3.9).<br />

Each command is shown with its full name, short name (if one exists), optional parameters (in<br />

brackets), parameter range (if appropriate), and detailed information.<br />

Temp [d.d]<br />

• Default parameter: 20.0 (°C)<br />

• Parameter range: -5.0 to 60.0 (°C)<br />

• Without parameter: returns its current setting.<br />

• When used with a valid input parameter, the user-specified temperature is set to this value.<br />

• This temperature value is used to calculate sound speed if TempMode is set to USER.<br />

• See the <strong>Argonaut</strong> Principles of Operation for the effect of sound speed on velocity data.<br />

Sal [d.d]<br />

• Default parameter: 34.5 (ppt)<br />

• Parameter range: 0.0 to 60.0 (ppt)<br />

• Without parameter: returns its current setting.<br />

• When used with a valid input parameter, the salinity is set to this value.<br />

• This salinity value is used to calculate sound speed.<br />

• See the <strong>Argonaut</strong> Principles of Operation for the effect of sound speed on velocity data.<br />

AvgInterval or AI [d]<br />

• Default parameter: 120 (s)<br />

• Parameter range: 10 to 3600 (s; maximum is equivalent to 60 minutes)<br />

• Without parameter: returns its current setting (in seconds).<br />

• When used with a valid integer parameter, the averaging interval is set to this value.<br />

SampleInterval or SI [d]<br />

• Default parameter: 1200 (s; 20 minutes)<br />

• Parameter range: 10 to 43200 (s; maximum is equivalent to 12 hours)<br />

• Without parameter: returns its current setting (in seconds).<br />

• When used with a valid integer parameter, the sample interval is set to this value.<br />

• Sample interval is the time between the start of successive samples (§1.2 and §1.3).<br />

• If AvgInterval > SampleInterval, then AvgInterval takes precedence.<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)