Appendix 4. Analog Output Option for <strong>Argonaut</strong> SL/<strong>XR</strong> Systems<br />

A4.1. Introduction<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)<br />

<strong>SonTek</strong>/<strong>YSI</strong><br />

Beginning with <strong>Argonaut</strong> firmware version 7.9 (April, 2001), the ability to output data in an<br />

analog format was added to <strong>Argonaut</strong> SL and <strong>XR</strong> systems. This appendix describes the Analog<br />

Output (AO) option and operational considerations.<br />

A4.2. Overview of the <strong>Argonaut</strong> SL/<strong>XR</strong> Analog Output Option<br />

The AO implementation in the <strong>Argonaut</strong> can store and output up to eight different AO channels,<br />

each configured individually. You can select a different output variable for each channel. Nine<br />

output variables are available – flow, velocity magnitude, level, X-velocity, Y-velocity, average<br />

amplitude, pressure, temperature, and cell end location. When you are using more than one<br />

serial-to-AO converter (described in §0 and hereafter referred to as Converter), each Converter<br />

must be configured for a unique address (1 through 8).<br />

The following commands have been added to support the AO option.<br />

AnalogOutputType or AOT [DISABLED|CURRENT|VOLTAGE]<br />

• Sets the type of output to either CURRENT (4-20 mA) or VOLTAGE.<br />

• When set to DISABLED, the analog output is not performed.<br />

SetupAnalogOutput or SAO [Chan OutParam MinVal MaxVal] | [Chan RESET]<br />

• Sets or displays the AO processing parameters.<br />

• Chan is the channel number (1 through 8) of the <strong>Argonaut</strong>.<br />

• OutParam is the desired analog variable you wish to output. The acceptable parameters are<br />

VELX (current velocity of the X component), VELY (current velocity of the Y component),<br />

VELMAG (velocity magnitude), AVGAMP (average signal amplitude), TEMP (water temperature),<br />

PRESS (pressure), LEVEL (of water above <strong>Argonaut</strong>; only available with verticalbeam<br />

systems), FLOW (not yet implemented), and CELLEND (end location of cell; only<br />

available in systems with a pressure sensor).<br />

• MinVal and MaxVal are the minimum and maximum parameter limits. The reason for<br />

specifying the limit values is to allow you to obtain more accurate readings for your<br />

particular environmental conditions. For example, the velocity magnitude range for the<br />

<strong>Argonaut</strong> current meter is from 0 to 1000 cm/s. These values correspond to the<br />

Converter’s limits of 4 and 20 mA, respectively. However, if the expected velocity<br />

magnitude is >125 and