3.2.1. Command Mode<br />

8<br />

• The <strong>Argonaut</strong> can send and receive commands related to all aspects of instrument<br />

operation.<br />

• You can enter the command mode from any other mode by sending a BREAK (§3.4).<br />

• You can enter the command mode from the data acquisition mode using the run-time<br />

command “+++” (§3.15).<br />

• You can put the <strong>Argonaut</strong> into any other mode only from the command mode.<br />

• To enter the data acquisition mode from the command mode, use the command Start.<br />

• To enter the deployment mode from the command mode, use the command Deploy.<br />

• To enter the SDI-12 interface mode from the command mode, use the command sdi12.<br />

• To enter the sleep mode from the command mode, use the command PowerOff.<br />

• If the <strong>Argonaut</strong> is left idle in the command mode for more than five minutes, it will enter<br />

the sleep mode to conserve power.<br />

3.2.2. Data Acquisition Mode<br />

• Data acquisition mode is used for real-time data collection for which you are typically<br />

connected to an external power supply and computer or data logger.<br />

• Data acquisition mode is entered from command mode with the command Start. After<br />

the Start command, the <strong>Argonaut</strong> takes a few seconds to initialize and then begins data<br />

collection.<br />

• In data acquisition mode, the <strong>Argonaut</strong> ignores the deployment parameters StartDate and<br />

StartTime.<br />

• If the internal recorder has been enabled (Recorder ON), data are output both over the<br />

serial port and to the internal recorder. If the recorder has been disabled (Recorder OFF),<br />

data are sent only to the serial port.<br />

• You can exit the data acquisition mode and return to the command mode by sending a<br />

BREAK (§3.4) or by using the run-time command “+++”.<br />

• While in data acquisition mode, the <strong>Argonaut</strong> can enter a low-power state between pings<br />

and between samples. This state is similar to, although not the same as, the sleep mode.<br />

The <strong>Argonaut</strong> will enter the low-power state if the system command AutoSleep is set to<br />

ON (the default setting). See §3.7 regarding the AutoSleep command and §3.15 regarding<br />

the effects of the power saving state on the run-time commands.<br />

3.2.3. Deployment Mode<br />

• Deployment mode is used for autonomous data collection (internal recording, typically<br />

with battery power).<br />

• Deployment mode is entered from command mode using the command Deploy.<br />

• In deployment mode, the instrument starts data collection at the date and time specified by<br />

StartDate and StartTime. If the current date and time are after the specified start date<br />

and time, the system begins data collection immediately.<br />

• In deployment mode, data are always stored to the internal recorder regardless of the<br />

Recorder ON/OFF parameter.<br />

• In deployment mode, the <strong>Argonaut</strong> will always enter the low-power state between pings<br />

and between samples regardless of the AutoSleep parameter (AutoSleep is forced ON).<br />

• Because of the power saving state, the run-time commands (§3.15) cannot be used.<br />

• You can exit the deployment mode and enter the command mode by sending a BREAK.<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)