SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>SonTek</strong>/<strong>YSI</strong><br />
3.2.1. Command Mode<br />
8<br />
• The <strong>Argonaut</strong> can send and receive commands related to all aspects of instrument<br />
operation.<br />
• You can enter the command mode from any other mode by sending a BREAK (§3.4).<br />
• You can enter the command mode from the data acquisition mode using the run-time<br />
command “+++” (§3.15).<br />
• You can put the <strong>Argonaut</strong> into any other mode only from the command mode.<br />
• To enter the data acquisition mode from the command mode, use the command Start.<br />
• To enter the deployment mode from the command mode, use the command Deploy.<br />
• To enter the SDI-12 interface mode from the command mode, use the command sdi12.<br />
• To enter the sleep mode from the command mode, use the command PowerOff.<br />
• If the <strong>Argonaut</strong> is left idle in the command mode for more than five minutes, it will enter<br />
the sleep mode to conserve power.<br />
3.2.2. Data Acquisition Mode<br />
• Data acquisition mode is used for real-time data collection for which you are typically<br />
connected to an external power supply and computer or data logger.<br />
• Data acquisition mode is entered from command mode with the command Start. After<br />
the Start command, the <strong>Argonaut</strong> takes a few seconds to initialize and then begins data<br />
collection.<br />
• In data acquisition mode, the <strong>Argonaut</strong> ignores the deployment parameters StartDate and<br />
StartTime.<br />
• If the internal recorder has been enabled (Recorder ON), data are output both over the<br />
serial port and to the internal recorder. If the recorder has been disabled (Recorder OFF),<br />
data are sent only to the serial port.<br />
• You can exit the data acquisition mode and return to the command mode by sending a<br />
BREAK (§3.4) or by using the run-time command “+++”.<br />
• While in data acquisition mode, the <strong>Argonaut</strong> can enter a low-power state between pings<br />
and between samples. This state is similar to, although not the same as, the sleep mode.<br />
The <strong>Argonaut</strong> will enter the low-power state if the system command AutoSleep is set to<br />
ON (the default setting). See §3.7 regarding the AutoSleep command and §3.15 regarding<br />
the effects of the power saving state on the run-time commands.<br />
3.2.3. Deployment Mode<br />
• Deployment mode is used for autonomous data collection (internal recording, typically<br />
with battery power).<br />
• Deployment mode is entered from command mode using the command Deploy.<br />
• In deployment mode, the instrument starts data collection at the date and time specified by<br />
StartDate and StartTime. If the current date and time are after the specified start date<br />
and time, the system begins data collection immediately.<br />
• In deployment mode, data are always stored to the internal recorder regardless of the<br />
Recorder ON/OFF parameter.<br />
• In deployment mode, the <strong>Argonaut</strong> will always enter the low-power state between pings<br />
and between samples regardless of the AutoSleep parameter (AutoSleep is forced ON).<br />
• Because of the power saving state, the run-time commands (§3.15) cannot be used.<br />
• You can exit the deployment mode and enter the command mode by sending a BREAK.<br />
<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)