Section 5. <strong>Argonaut</strong> Hardware<br />

5.1. <strong>Argonaut</strong> Hardware Configuration Settings<br />

<strong>SonTek</strong>/<strong>YSI</strong><br />

The hardware configuration of the different <strong>Argonaut</strong> systems depends on the requirements of the<br />

application. <strong>Argonaut</strong> hardware configuration settings can be displayed using the Show Conf<br />

command from the direct command interface (§3.14). A sample output of this command for an<br />

<strong>Argonaut</strong> SL is given below.<br />

>show conf<br />

HARDWARE CONFIGURATION PARAMETERS<br />

---------------------------------<br />

System Type -------------------- SL<br />

Sensor serial # ---------------- E15<br />

Sensor frequency - (kHz) ------- 1500<br />

Number of beams ---------------- 2<br />

Beam Geometry ------------------ 2_BEAMS<br />

Slant angle - (deg) ------------ 25.0<br />

Orientation -------------------- SIDE<br />

Compass installed ------------- NO<br />

Recorder installed ------------- YES<br />

Temperature sensor ------------- YES<br />

Pressure sensor ------------- NO<br />

Ctd sensor ------------- NO<br />

PressOffset - (dbar) ---------- 0.000000<br />

PressScale -- (dbar/count) ---- 0.000000<br />

PressScale_2 - (pdbar/count^2) - 0<br />

A brief description of each parameter is given below.<br />

• System type: Set at the factory for MD, <strong>XR</strong>, or SL.<br />

• Sensor serial number: The <strong>Argonaut</strong> serial number is stamped on the transducer head.<br />

• Sensor frequency: Specifies the acoustic frequency of the <strong>Argonaut</strong> in kHz (3000 or 1500).<br />

• Number of beams / Beam geometry / Slant angle: The <strong>Argonaut</strong> MD and <strong>XR</strong> use three<br />

beams for vertical operation. The <strong>Argonaut</strong> SL uses two beams for horizontal current<br />

measurements. The mounting angle of the <strong>Argonaut</strong> transducers is 25° for the <strong>Argonaut</strong> <strong>XR</strong><br />

and SL, 45° for the <strong>Argonaut</strong> MD. See <strong>Argonaut</strong> Principles of Operation for details on the<br />

exact beam geometry and mounting angle used by each system.<br />

• Orientation: Measurement orientation of the system for up-looking, down-looking, or sidelooking<br />

operation. Note that this setting must match the compass installation for correct<br />

operation (§4.2). Also, note that the orientation will affect the definition of the XYZ<br />

coordinate system (§6.4).<br />

• Compass / Recorder / Temperature / Pressure sensor: These settings specify which optional<br />

sensors have been installed. All systems include the recorder and temperature standard. The<br />

<strong>Argonaut</strong> MD and <strong>XR</strong> include a compass standard. The <strong>Argonaut</strong> <strong>XR</strong> includes a pressure<br />

sensor standard.<br />

• PressOffset / PressScale / PressScale_2: Factory calibration parameters for the pressure<br />

sensor (§6.6).<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001) 47