01.06.2013 Views

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Argonaut</strong> SL / SW / <strong>XR</strong> Firmware Version 11.7 (December 7, 2006)<br />

<strong>SonTek</strong>/<strong>YSI</strong> Firmware Release Notes<br />

6837 Nancy Ridge Dr., Suite A • San Diego, CA 92121 • Telephone (858) 546-8327 • Fax (858) 546-8150 • Internet: www.sontek.com<br />

<strong>Argonaut</strong> SL/SW/<strong>XR</strong> Firmware Versions 11.6 and 11.7<br />

View<strong>Argonaut</strong> Software Version 3.43<br />

<strong>SonTek</strong>/<strong>YSI</strong> has released <strong>Argonaut</strong> SL/SW/<strong>XR</strong> firmware versions 11.6 and 11.7. To take advantage<br />

of the new features, you should also update to View<strong>Argonaut</strong> software version 3.43 or later.<br />

If you have any questions about these new features, please contact <strong>SonTek</strong>/<strong>YSI</strong>.<br />

NOTE: This document mainly describes firmware version 11.6. Version 11.7 was released December<br />

7, 2006 to correct a minor diagnostic issue discovered shortly after the release of v11.6<br />

• Section 1 describes changes and new features available in versions 11.6 and 11.7.<br />

• Section 2 gives information for updating your system to firmware version 11.7.<br />

o Note: All production systems shipped on or after December 7, 2006 have version 11.7<br />

already installed and do not require you to update the firmware.<br />

• Section 3 gives contact information for <strong>SonTek</strong>/<strong>YSI</strong>.<br />

1. Changes and New Features in Version 11.6 and 11.7<br />

1.1. Total volume criteria<br />

Total volume calculations allow the <strong>Argonaut</strong> to determine the amount of water that has passed<br />

the system over a given span of time; new criteria allow the modification of this calculation.<br />

• Total volume is the cumulative sum of flow rate multiplied by time. This represents the<br />

amount of water that has passed the system in a given amount of time.<br />

• By default, total volume includes all flow past the system regardless of the magnitude or direction<br />

of the flow.<br />

• New criteria modify the calculations to accumulate flow into total volume.<br />

o Total volume is only accumulated if the flow rate meets certain criteria:<br />

Flow ≥ a user-specified threshold<br />

Flow ≤ a user-specified threshold<br />

Magnitude (absolute value) of flow ≥ a user-specified threshold<br />

o Total volume is only accumulated if the velocity meets certain criteria:<br />

Velocity ≥ a user-specified threshold<br />

Velocity ≤ a user-specified threshold<br />

Magnitude (absolute value) of velocity ≥ a user-specified threshold<br />

• These conditions are provided to calculate a modified total volume value (e.g., only accumulate<br />

flow in a certain direction) or to avoid the accumulation of small, residual flow values<br />

in stagnant water.<br />

1.2. Change water level output to stage<br />

All real-time data from the vertical beam (SL and SW systems) are now output as “stage”<br />

rather than “water level”.<br />

• Water level is the height above the water surface above the top of the system.<br />

• Stage is water level plus the user-specified instrument elevation. This value thus reports water<br />

depth relative to a user-specified datum.<br />

• All real-time data outputs (RS232, SDI-12, analog output, and Modbus) now output stage<br />

rather than water level.<br />

1

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!