01.06.2013 Views

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>SonTek</strong>/<strong>YSI</strong><br />

SONTEK Output Data Format<br />

Standard (Single Averaged-Cell) Data: To request and retrieve a standard (i.e., single averagedcell)<br />

sample in SDI-12 SONTEK mode, issue these commands:<br />

Command Response Description _<br />

“aM!” or “aM0!” Initiate measurement request<br />

“aD0!” a±T.TT±P.PPP±L.LLL±CC.C Retrieve 1 st and 2 nd buffers of standard (single<br />

“aD1!” a±X.XX±Y.YY±Z.ZZ±S. averaged-cell) data<br />

102<br />

Where: (See below for data format description legend.)<br />

Profiling Cell Data: To request and retrieve profiling cell sample data in SDI-12 SONTEK mode,<br />

issue the following commands. Profiling cell data are only available in systems that have the<br />

optional Profiling Mode feature installed.<br />

Command Response Description _<br />

“aM!” or “aM0!” Initiate measurement request<br />

“aD0!” a±T.TT±P.PPP±L.LLL±CC.C Retrieve 1 st and 2 nd buffers of standard (single<br />

“aD1!” a±X.XX±Y.YY±Z.ZZ±S. averaged-cell) data<br />

“aMn!” Request profiling cell (n = cell number 1-5) ††<br />

“aD0!” a±x.xx±y.yy±z.zz±s. Retrieve data for profiling cell n<br />

††<br />

Note: When issuing the “aMn!” command, the data that is sent will be the data that was stored<br />

in the buffer when the “aM!” command was issued. That is, subsequently issuing the same<br />

“aMn!” command will not display new data unless another “aM!” command is issued before<br />

issuing the “aMn!” command.<br />

Where:<br />

Data Format<br />

Standard (Single-Averaged) Cell:<br />

Metric English<br />

T.TT = Mean Temperature °C °F<br />

P.PPP = Mean Pressure decibar psi<br />

L.LLL = Level cm ft<br />

CC.C = Cell End m ft<br />

X.XX = Velocity component 1* (beam 1/X/East) cm/s ft/s<br />

Y.YY = Velocity component 2* (beam 2/Y/North) cm/s ft/s<br />

Z.ZZ = Velocity component 3* (beam 3/Z/Up) [<strong>XR</strong> only] cm/s ft/s<br />

or Velocity Magnitude [SL only] cm/s ft/s<br />

S. = Mean Signal Strength counts counts<br />

Individual Profiling Cell (must have Profiling Mode option installed):<br />

x.xx = Velocity component 1 † (beam 1/X/East) cm/s ft/s<br />

y.yy = Velocity component 2 † (beam 2/Y/North) cm/s ft/s<br />

z.zz = Velocity component 3 † (beam 3/Z/Up) [<strong>XR</strong> only] cm/s ft/s<br />

or Velocity Magnitude † [SL only] cm/s ft/s<br />

s. = Mean Signal Strength † counts counts<br />

Notes: * Cell parameters determined by CellBegin (CB) and CellEnd (CE).<br />

†<br />

Cell parameters determined by CellSize (CS), BlankDistance (BD), and the cell<br />

number (1 through 5) selected in the “aMn!” command.<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!