Appendix 3. Vertical Beam Support for <strong>Argonaut</strong> SL Systems<br />

A3.1. Introduction<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)<br />

<strong>SonTek</strong>/<strong>YSI</strong><br />

Beginning with <strong>Argonaut</strong> firmware version 5.7 (April, 2000), support was added for <strong>Argonaut</strong><br />

SL systems that have the optional vertical beam installed. Differences between standard<br />

<strong>Argonaut</strong> commands and data outputs are described in this appendix.<br />

A3.2. Additional Commands for Vertical Beam <strong>Argonaut</strong> SL Systems<br />

The following commands have been added to support vertical beam operation in the SL.<br />

VerticalBeam Set [YES|NO]<br />

• Enables or disables the use of the vertical beam.<br />

• When set to YES, processing of vertical beam data requires additional CPU resources. As<br />

such, the minimum value for the averaging interval is 10 seconds.<br />

Level [CONT]<br />

• This command outputs the current water level value in mm.<br />

• When used with the CONT parameter, the water level value is continually output until the<br />

spacebar is pressed.<br />

A3.3. Real-Time Data Format for Vertical Beam <strong>Argonaut</strong> SL Systems<br />

The following changes apply to the output data stream for vertical beam systems. See §3.16 for a<br />

full description of the <strong>Argonaut</strong>’s output data format.<br />

• When DataFormat = LONG, water level is reported instead of the third velocity component<br />

(V3/Vz/Vup).<br />

• When DataFormat = SHORT, water level is reported instead of the pressure field. Note: If<br />

you need to report pressure data, either disable the vertical beam (VerticalBeam Set NO)<br />

or use DataFormat LONG.<br />

• The water level measurement unit depends on the setting of OutFormat: ASCII = mm,<br />

Metric = cm, English = ft, Binary = mm.<br />

• When no water level is detected, a value of –1 is reported.<br />

111