Section 10. Recorder Data Extraction with SONREC (DOS)<br />

This program is used to extract data from the <strong>Argonaut</strong> internal recorder. To learn about its<br />

operation, type SONREC at the DOS prompt.<br />

<strong>SonTek</strong>/<strong>YSI</strong><br />

IMPORTANT: This program should be run from a dedicated DOS environment, not from a<br />

DOS prompt within Windows. Memory resident features of Windows interfere with serial port<br />

operation. Running this program from within Windows will, at the very least, slow down data<br />

retrieval, and can potentially cause the retrieval to fail.<br />

To run from the DOS prompt, use one of the following commands.<br />

SONREC <br />

SONREC <br />

SONREC <br />

where<br />

is the serial port number to which the <strong>Argonaut</strong> is connected (1 for COM1 or<br />

2 for COM2).<br />

is the communication baud rate setting of the <strong>Argonaut</strong>; if no value is given, the<br />

program assumes the default setting of 9600 baud. Acceptable baud rate values are 1200,<br />

2400, 4800, 9600, 19200, 38400, 57600, and 115200.<br />

is the baud rate at which data files will be retrieved from the recorder; if no<br />

value is given, it uses the default setting of 115200 baud. Acceptable extraction rate values<br />

are 1200, 2400, 4800, 9600, 19200, 38400, 57600, and 115200.<br />

For example, the command “SONREC 1 9600 38400” will establish communication with the<br />

<strong>Argonaut</strong> on COM1 at 9600 baud and download data files using 38400 baud.<br />

For short cables (less than 30 m), the default extraction rate of 115200 baud provides the fastest<br />

downloading time. For longer cables, a slower baud rate may be needed to ensure reliable<br />

communication. If SONREC encounters a large number of communication errors while retrieving<br />

data, it will terminate the retrieval. Note that some computers (particularly older laptop<br />

computers) do not operate reliably at high baud rates and may not be able to work using the<br />

default setting of 115200. If you encounter problems with SONREC, run the program using a lower<br />

extraction rate.<br />

The table below shows the typical download rates achieved with different baud rate settings.<br />

Extraction rates slower than 9600 baud are rarely needed and are not shown here. Actual retrieval<br />

rates may vary depending upon the environment and the type of computer used.<br />

SONREC Data Retrieval Rates<br />

Extraction rate Effective retrieval rate Download time for 1 MB<br />

115200 4800 bytes per second 3.6 minutes<br />

57600 3400 bytes per second 5.1 minutes<br />

38400 2600 bytes per second 6.7 minutes<br />

19200 1500 bytes per second 11.5 minutes<br />

9600 850 bytes per second 20.5 minutes<br />

<strong>Argonaut</strong> Software <strong>Manual</strong> (March 1, 2001) 67