4. Additional System Commands<br />

This section describes the system commands used for external pressure sensor operation.<br />

4.1. ADPs and Hydras<br />

The following commands apply to ADP and Hydra (ADVOcean) systems that have an external<br />

pressure sensor installed.<br />

ExtPressureSensorInstalled Set [NONE|PAROS|DRUCK|PAROSFREQ]<br />

• Selects or disables the external pressure sensor type.<br />

• Set at the factory. Does not need to be changed except in some special cases.<br />

• The external pressure sensor option limits the maximum sampling rate of an ADV to 4 Hz<br />

for the PAROS and to 10 Hz for a DRUCK or PAROSFREQ. To collect data at faster rates, you<br />

must disable the external pressure sensor option.<br />

PressFreqOffset Set [d]<br />

• Default parameters (in kHz): 0 (for DRUCK); 21 (for PAROSFREQ)<br />

• Sets/displays frequency offset (in kHz).<br />

• This parameter is set at the factory and should not be changed by the user.<br />

PAROS CONT or DRUCK CONT or PAROSFREQ CONT<br />

• This command provides a useful tool to verify and troubleshoot the frequency sensor.<br />

• For PAROS sensors:<br />

o This command continually outputs pressure reading in dbar.<br />

o If the pressure reading is zero, the connection between the sensor and the processing<br />

electronics may be damaged (or open).<br />

• For DRUCK and PAROSFREQ sensors:<br />

o This command outputs a sample consisting of converted pressure (dbar), temperature<br />

(°C), and raw frequency counts (Hz) in the following format:<br />

4<br />

Druck Press = 10.03192 Temp = 22.11 Freq = 8055.674<br />

o The DRUCK frequency range (in air) is 7 to 10 kHz; the PAROSFREQ is 35 to 38 kHz.<br />

o If the frequency reading is zero, or is outside the range, the connection between the sensor<br />

and processing electronics may be damaged (or open).<br />

o If the frequency reading is reasonable, but the corresponding pressure is not, this indicates<br />

that the appropriate calibration is not loaded into the processor. Section 5 describes<br />

how to load the DRUCK calibration.<br />