<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)<br />
<strong>SonTek</strong>/<strong>YSI</strong><br />
The same communications cable is used for both<br />
SDI-12 and RS-232 operations. A white, 2-pin<br />
switch plug is used to change the hardware between<br />
the two modes (Figure A2.1); a software<br />
command also needs to be sent (see §A2.3).<br />
When both ends of the plug are connected (the<br />
plug is shorted), RS-232 communication is en- Figure A2.2. Switching hardware between<br />
abled. When both ends of the plug are separated SDI-12 and RS-232 modes.<br />
(the plug is open), SDI-12 communication mode<br />
is enabled (Figure A2.2). Suggested connection sequence is as follows (Figure A2.3):<br />
• Attach 8-pin underwater connector on the communications cable to the <strong>Argonaut</strong>.<br />
• Connect serial DB-9 connector to a serial port on your computer.<br />
• Connect SDI-12 pigtail to the data-logger.<br />
Most data-loggers have +12V power output that can be used to power the <strong>Argonaut</strong> through the<br />
SDI-12 cable (red lead) directly.<br />
When using the supplied AC-to-DC power supply, or when operating the <strong>Argonaut</strong> with a power<br />
source that exceeds 16 VDC, or when operating a data logger that does not provide power to external<br />
sensors, the <strong>Argonaut</strong> can be powered using the coaxial power connector (up to 26 VDC).<br />
A2.2.1. DC Power<br />
The <strong>Argonaut</strong> requires 6-16 VDC input power (12.0 VDC recommended). Normally the <strong>Argonaut</strong><br />
is a low power device, however, when it transmits acoustic pulses it requires up to 200 mA<br />
for short bursts. Make sure your wiring and battery are capable of supplying sufficient current.<br />
Datalogger<br />
Red- power (6 to 16 V)<br />
White- data (SDI-12)<br />
Black- ground<br />
Serial connector DB-9<br />
(to computer)<br />
SDI-12 cable<br />
Coaxial power connector (10 to 26 V)<br />
(to optional external source)<br />
Figure A2.3. Connecting an <strong>Argonaut</strong>-SL to a data-logger.<br />
<strong>Argonaut</strong><br />
8-pin underwater<br />
connector<br />