<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)<br />

<strong>SonTek</strong>/<strong>YSI</strong><br />

The same communications cable is used for both<br />

SDI-12 and RS-232 operations. A white, 2-pin<br />

switch plug is used to change the hardware between<br />

the two modes (Figure A2.1); a software<br />

command also needs to be sent (see §A2.3).<br />

When both ends of the plug are connected (the<br />

plug is shorted), RS-232 communication is en- Figure A2.2. Switching hardware between<br />

abled. When both ends of the plug are separated SDI-12 and RS-232 modes.<br />

(the plug is open), SDI-12 communication mode<br />

is enabled (Figure A2.2). Suggested connection sequence is as follows (Figure A2.3):<br />

• Attach 8-pin underwater connector on the communications cable to the <strong>Argonaut</strong>.<br />

• Connect serial DB-9 connector to a serial port on your computer.<br />

• Connect SDI-12 pigtail to the data-logger.<br />

Most data-loggers have +12V power output that can be used to power the <strong>Argonaut</strong> through the<br />

SDI-12 cable (red lead) directly.<br />

When using the supplied AC-to-DC power supply, or when operating the <strong>Argonaut</strong> with a power<br />

source that exceeds 16 VDC, or when operating a data logger that does not provide power to external<br />

sensors, the <strong>Argonaut</strong> can be powered using the coaxial power connector (up to 26 VDC).<br />

A2.2.1. DC Power<br />

The <strong>Argonaut</strong> requires 6-16 VDC input power (12.0 VDC recommended). Normally the <strong>Argonaut</strong><br />

is a low power device, however, when it transmits acoustic pulses it requires up to 200 mA<br />

for short bursts. Make sure your wiring and battery are capable of supplying sufficient current.<br />

Datalogger<br />

Red- power (6 to 16 V)<br />

White- data (SDI-12)<br />

Black- ground<br />

Serial connector DB-9<br />

(to computer)<br />

SDI-12 cable<br />

Coaxial power connector (10 to 26 V)<br />

(to optional external source)<br />

Figure A2.3. Connecting an <strong>Argonaut</strong>-SL to a data-logger.<br />

<strong>Argonaut</strong><br />

8-pin underwater<br />

connector<br />

95