5. Additional Software<br />

This section describes additional software required to use external frequency pressure sensors<br />

(DRUCK/PAROSFREQ) with <strong>SonTek</strong> instruments.<br />

5.1. Downloading and Extracting Druck Calibration: LDDRUCK/GTDRUCK<br />

The LDDRUCK.EXE program loads the DRUCK/PAROSFREQ calibration file into the instrument’s<br />

processor. The command syntax is:<br />

LDDRUCK [DruckCalFile] [-COM port] [-Baud Rate] [-Address]<br />

where<br />

6<br />

• DruckCalFile is the name of the calibration file (no extension).<br />

• -COM port is the serial port to which the system is connected (1=COM1; 2= COM2).<br />

• -Baud Rate is the communication baud rate setting of the system. If no value is given, the<br />

program assumes the default setting of 9600 baud. Acceptable baud rate settings are 1200,<br />

2400, 4800, 9600, 19200, 38400, 57600, and 115200.<br />

• -Address is the address for a RS-485 system.<br />

Example: LDDRUCK P84567 –p1 –b19200 –a2 will download the file P84567.DRK onto the<br />

system at address 2 connected to COM port 1 at 19200 baud.<br />

The GTDRUCK.EXE program gets (downloads) the DRUCK/PAROSFREQ calibration file from the instrument’s<br />

processor to the user’s PC. The command syntax is:<br />

GTDRUCK [MyCalFile] [-COM port] [-Baud Rate] [-Address]<br />

where<br />

• MyCalFile is the name of the calibration file (no extension).<br />

• -COM port is the serial port to which the system is connected (1=COM1; 2= COM2).<br />

• -Baud Rate is the communication baud rate setting of the system. If no value is given, the<br />

program assumes the default setting of 9600 baud. Acceptable baud rate settings are 1200,<br />

2400, 4800, 9600, 19200, 38400, 57600, and 115200.<br />

• -Address is the address for a RS-485 system.<br />

Example: GTDRUCK B199 –p2 –b19200 will retrieve the calibration data from the system<br />

connected to COM port 2 at 19200 baud and store the data into file B199.DRK.<br />

Using Frequency Pressure Sensors with <strong>SonTek</strong>/<strong>YSI</strong> Systems (June 2001)