SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Where:<br />
Data Format Metric English<br />
x.xx = Velocity component 1 † (beam 1/X/East) cm/s ft/s<br />
y.yy = Velocity component 2 † (beam 2/Y/North) cm/s ft/s<br />
z.zz = Velocity component 3 † (beam 3/Z/Up) [<strong>XR</strong> only] cm/s ft/s<br />
or Velocity Magnitude † [SL only] cm/s ft/s<br />
s. = Mean Signal Strength † counts counts<br />
† Note: Cell parameters determined by CellSize (CS), BlankDistance (BD), Ncells (NC), and<br />
the cell number selected in the “aMc!” command (cell number 4 in this example).<br />
Example of “aD0!” through “aDx!” commands following “aC!” command, where<br />
sdi12format=SONTEK, PM=YES, and NC=3:<br />
Previous Command Response Description<br />
“aC!” “a06020” See §A2.5.2<br />
Subsequent Command Response Description<br />
“aD0!” a±T.TT±P.PPP±L.LLL±CC.C See §A2.5.3<br />
“aD1!” a±X.XX±Y.YY±Z.ZZZ±S.<br />
“aD2!” a±x.xx±y.yy±z.zz±s. [for Cell 1]<br />
“aD3!” a±x.xx±y.yy±z.zz±s. [for Cell 2]<br />
“aD4!” a±x.xx±y.yy±z.zz±s. [for Cell 3]<br />
Where:<br />
Data Format Metric English<br />
T.TT = Mean Temperature °C °F<br />
P.PPP = Mean Pressure decibar psi<br />
L.LLL = Level cm ft<br />
CC.C = Cell End m ft<br />
X.XX = Velocity component 1* (beam 1/X/East) cm/s ft/s<br />
Y.YY = Velocity component 2* (beam 2/Y/North) cm/s ft/s<br />
Z.ZZ = Velocity component 3* (beam 3/Z/Up) [<strong>XR</strong> only] cm/s ft/s<br />
or Velocity Magnitude [SL only] cm/s ft/s<br />
S. = Mean Signal Strength counts counts<br />
x.xx = Velocity component 1 † (beam 1/X/East) cm/s ft/s<br />
y.yy = Velocity component 2 † (beam 2/Y/North) cm/s ft/s<br />
z.zz = Velocity component 3 † (beam 3/Z/Up) [<strong>XR</strong> only] cm/s ft/s<br />
or Velocity Magnitude [SL only] cm/s ft/s<br />
s. = Mean Signal Strength † counts counts<br />
Notes: *Cell parameters determined by CellBegin (CB) and CellEnd (CE).<br />
† Cell parameters determined by CellSize (CS), BlankDistance (BD), Ncells (NC), and<br />
the appropriate cell number (cells 1 through 3 in this example).<br />
<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)<br />
<strong>SonTek</strong>/<strong>YSI</strong><br />
107