Where:<br />

Data Format Metric English<br />

x.xx = Velocity component 1 † (beam 1/X/East) cm/s ft/s<br />

y.yy = Velocity component 2 † (beam 2/Y/North) cm/s ft/s<br />

z.zz = Velocity component 3 † (beam 3/Z/Up) [<strong>XR</strong> only] cm/s ft/s<br />

or Velocity Magnitude † [SL only] cm/s ft/s<br />

s. = Mean Signal Strength † counts counts<br />

† Note: Cell parameters determined by CellSize (CS), BlankDistance (BD), Ncells (NC), and<br />

the cell number selected in the “aMc!” command (cell number 4 in this example).<br />

Example of “aD0!” through “aDx!” commands following “aC!” command, where<br />

sdi12format=SONTEK, PM=YES, and NC=3:<br />

Previous Command Response Description<br />

“aC!” “a06020” See §A2.5.2<br />

Subsequent Command Response Description<br />

“aD0!” a±T.TT±P.PPP±L.LLL±CC.C See §A2.5.3<br />

“aD1!” a±X.XX±Y.YY±Z.ZZZ±S.<br />

“aD2!” a±x.xx±y.yy±z.zz±s. [for Cell 1]<br />

“aD3!” a±x.xx±y.yy±z.zz±s. [for Cell 2]<br />

“aD4!” a±x.xx±y.yy±z.zz±s. [for Cell 3]<br />

Where:<br />

Data Format Metric English<br />

T.TT = Mean Temperature °C °F<br />

P.PPP = Mean Pressure decibar psi<br />

L.LLL = Level cm ft<br />

CC.C = Cell End m ft<br />

X.XX = Velocity component 1* (beam 1/X/East) cm/s ft/s<br />

Y.YY = Velocity component 2* (beam 2/Y/North) cm/s ft/s<br />

Z.ZZ = Velocity component 3* (beam 3/Z/Up) [<strong>XR</strong> only] cm/s ft/s<br />

or Velocity Magnitude [SL only] cm/s ft/s<br />

S. = Mean Signal Strength counts counts<br />

x.xx = Velocity component 1 † (beam 1/X/East) cm/s ft/s<br />

y.yy = Velocity component 2 † (beam 2/Y/North) cm/s ft/s<br />

z.zz = Velocity component 3 † (beam 3/Z/Up) [<strong>XR</strong> only] cm/s ft/s<br />

or Velocity Magnitude [SL only] cm/s ft/s<br />

s. = Mean Signal Strength † counts counts<br />

Notes: *Cell parameters determined by CellBegin (CB) and CellEnd (CE).<br />

† Cell parameters determined by CellSize (CS), BlankDistance (BD), Ncells (NC), and<br />

the appropriate cell number (cells 1 through 3 in this example).<br />