<strong>SonTek</strong>/<strong>YSI</strong><br />

A2.2. Connecting the <strong>Argonaut</strong> to a Data-Logger<br />

To facilitate operation of an <strong>Argonaut</strong> with external SDI-12 data storing systems, the <strong>Argonaut</strong><br />

power/communication cable has been modified to include a separate termination to be used with<br />

SDI-12 capable data-loggers (Figure A2.1).<br />

The cable has an 8-pin underwater connector on one end and three separate cable terminations on<br />

the other end. The description of these termination cables follows.<br />

Note: Although the cable below represents the most common assembly, the cable for your system<br />

may vary. See §A2.2.3 and Figures A2.4 through A2.7 for information on the most common<br />

cable assemblies.<br />

94<br />

• SDI-12 pigtail:<br />

Red ---------------------- Power to <strong>Argonaut</strong> (+6 to +16 VDC)<br />

White ---------------------- Data (SDI-12)<br />

Black ---------------------- Ground<br />

WARNING: The power line does not have over-voltage protection. Voltages<br />

above +16V on the pigtail will damage <strong>Argonaut</strong> electronics.<br />

• Coaxial power cable (for an optional AC/DC adaptor or external power above 16 VDC):<br />

Center pin ----------------- Power to <strong>Argonaut</strong> (+10 to +26 VDC)<br />

Jacket ----------------- Ground<br />

WARNING: The power line does not have over-voltage protection. Voltages<br />

above +26V on the coaxial power cable will damage <strong>Argonaut</strong> electronics.<br />

• RS-232 cable (DB9):<br />

Pin 2 ------------------------ Data out<br />

Pin 3 ------------------------ Data in<br />

Pin 5 ------------------------ Ground<br />

Note: The remaining pins are not connected.<br />

Figure A2.1. <strong>Argonaut</strong> power/communications cable<br />