<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)<br />
<strong>SonTek</strong>/<strong>YSI</strong><br />
<strong>Argonaut</strong> SHORT Sample Format<br />
• When SHORT data format is used, the data output contains 14 columns. The following table<br />
shows the contents and units used with each column.<br />
• For the SHORT data format, reported signal strength is the mean of all working beams<br />
(three beams for <strong>Argonaut</strong> <strong>XR</strong> and MD; two beams for <strong>Argonaut</strong> SL).<br />
• Reported standard deviation is the defined based on the coordinate system:<br />
Beam Coordinates σV = √(σV1 2 + σV2 2 + σV3 2 )<br />
XYZ Coordinates σV = √(σVx 2 + σVy 2 )<br />
ENU Coordinate σV = √(σVE 2 + σVN 2 )<br />
<strong>Argonaut</strong> ASCII/METRIC/ENGLISH Data Output - SHORT Data Format<br />
Col Contents ASCII Metric English<br />
1 Sample time (start of averaging interval) - Year<br />
2 Sample time (start of averaging interval) - Month<br />
3 Sample time (start of averaging interval) - Day<br />
4 Sample time (start of averaging interval) - Hour<br />
5 Sample time (start of averaging interval) - Minute<br />
6 Sample time (start of averaging interval) - Second<br />
7 Velocity component 1 (Beam 1/X/East) 0.1 cm/s cm/s ft/s<br />
8 Velocity component 2 (Beam 2/Y/North) 0.1 cm/s cm/s ft/s<br />
9 Velocity component 3 (Beam 3/Z/Up) 0.1 cm/s cm/s ft/s<br />
10 Mean standard deviation (see above) 0.1 cm/s cm/s ft/s<br />
11 Mean signal strength (see above) count count count<br />
12 Mean temperature 0.01°C °C °F<br />
13<br />
Mean pressure – OR –<br />
Water level (SL systems with vertical beam; Appendix 2)<br />
count<br />
mm<br />
dbar<br />
cm<br />
PSI<br />
ft<br />
14 Input power level 0.2 V V V<br />
15+ Note: Additional fields may appear depending on installed<br />
options. See Section 8 for the data format of wave spectra,<br />
CTD, and other optional sensors.<br />
37