3.4.2. SDI-12 Deployment<br />

Setup Parameters<br />

Hardware Configuration Window<br />

Deployment Button<br />

Figure 5 – View<strong>Argonaut</strong> SDI-12 Deployment Setup Window<br />

Figure 5 shows the SDI-12 deployment setup window.<br />

• The main area lists setup parameters.<br />

• The bottom has the control buttons.<br />

The basic procedure for an SDI-12 deployment is below.<br />

<strong>SonTek</strong>/<strong>YSI</strong><br />

• Connect the <strong>Argonaut</strong> to the PC and the power supply to be used for deployment.<br />

• Start View<strong>Argonaut</strong>/Deployment and select SDI-12 Deployment.<br />

• Verify hardware configuration parameters using System Configuration Parameters.<br />

• Specify all setup parameters for your application. Note that the data format parameter is<br />

automatically set to SHORT.<br />

• Click Deploy to start data collection.<br />

• Disconnect the communication cable from the PC before turning the computer off.<br />

• Connect the SDI-12 cable to the data logger.<br />

• Install dummy plugs and other cables as needed. Deploy the system.<br />

<strong>Argonaut</strong> Software <strong>Manual</strong> (March 1, 2001) 15