3.7. System Commands<br />
System commands relate to general operation of the <strong>Argonaut</strong>.<br />
• Starting data collection<br />
• Displaying the serial number<br />
• Setting the clock<br />
• Data output and storage<br />
<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)<br />
<strong>SonTek</strong>/<strong>YSI</strong><br />
Each command is shown with its full name, short name (if one exists), optional parameters (in<br />
brackets), and appropriate detailed information.<br />
Start<br />
• Starts real-time data collection, putting the <strong>Argonaut</strong> into data acquisition mode (§3.2.2).<br />
• The system first saves all recently entered commands (i.e., does a SaveSetup).<br />
• Data collection begins immediately (StartDate and StartTime are ignored).<br />
• Burst sampling cannot be used (burst sampling parameters are ignored).<br />
• The system enters a low-power state between samples if AutoSleep is ON.<br />
• Run-time commands can only be used when AutoSleep is OFF.<br />
• Data will be stored to the recorder if enabled (Recorder ON).<br />
Deploy (do not confuse with Deployment command, §3.9)<br />
• Starts autonomous data collection, putting the <strong>Argonaut</strong> into deployment mode (§3.2.3).<br />
• The system first saves all recently entered commands (i.e., does a SaveSetup).<br />
• Data collection begins when the <strong>Argonaut</strong> clock reaches the specified StartDate and<br />
StartTime. If the current date/time is after the specified start date/time, data collection<br />
begins immediately.<br />
• <strong>Argonaut</strong> data are stored to the recorder regardless of the Recorder setting.<br />
• Instructs system to enter a low-power state between samples (AutoSleep is forced ON).<br />
SDI12 ON<br />
• The sdi12 ON command is used only with the SDI-12 serial interface.<br />
• See Appendix 2 for a description of the SDI-12 interface.<br />
SaveSetup<br />
• Saves all current parameter settings to internal memory (EEPROM).<br />
• This command must be sent before the system is shut down (or before a BREAK is sent);<br />
otherwise, any recently entered commands will be lost.<br />
• Executed automatically as part of the Start, Deploy, and sdi12 commands.<br />
Defaults or DEF<br />
• Sets all parameters (except baud rate) to the factory default values. The baud rate remains<br />
at the currently active value. To change the baud rate, see §5.4.<br />
• See the individual command descriptions for default values.<br />
Ver<br />
• Outputs the version number of the CPU firmware.<br />
15