3.7. System Commands<br />

System commands relate to general operation of the <strong>Argonaut</strong>.<br />

• Starting data collection<br />

• Displaying the serial number<br />

• Setting the clock<br />

• Data output and storage<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)<br />

<strong>SonTek</strong>/<strong>YSI</strong><br />

Each command is shown with its full name, short name (if one exists), optional parameters (in<br />

brackets), and appropriate detailed information.<br />

Start<br />

• Starts real-time data collection, putting the <strong>Argonaut</strong> into data acquisition mode (§3.2.2).<br />

• The system first saves all recently entered commands (i.e., does a SaveSetup).<br />

• Data collection begins immediately (StartDate and StartTime are ignored).<br />

• Burst sampling cannot be used (burst sampling parameters are ignored).<br />

• The system enters a low-power state between samples if AutoSleep is ON.<br />

• Run-time commands can only be used when AutoSleep is OFF.<br />

• Data will be stored to the recorder if enabled (Recorder ON).<br />

Deploy (do not confuse with Deployment command, §3.9)<br />

• Starts autonomous data collection, putting the <strong>Argonaut</strong> into deployment mode (§3.2.3).<br />

• The system first saves all recently entered commands (i.e., does a SaveSetup).<br />

• Data collection begins when the <strong>Argonaut</strong> clock reaches the specified StartDate and<br />

StartTime. If the current date/time is after the specified start date/time, data collection<br />

begins immediately.<br />

• <strong>Argonaut</strong> data are stored to the recorder regardless of the Recorder setting.<br />

• Instructs system to enter a low-power state between samples (AutoSleep is forced ON).<br />

SDI12 ON<br />

• The sdi12 ON command is used only with the SDI-12 serial interface.<br />

• See Appendix 2 for a description of the SDI-12 interface.<br />

SaveSetup<br />

• Saves all current parameter settings to internal memory (EEPROM).<br />

• This command must be sent before the system is shut down (or before a BREAK is sent);<br />

otherwise, any recently entered commands will be lost.<br />

• Executed automatically as part of the Start, Deploy, and sdi12 commands.<br />

Defaults or DEF<br />

• Sets all parameters (except baud rate) to the factory default values. The baud rate remains<br />

at the currently active value. To change the baud rate, see §5.4.<br />

• See the individual command descriptions for default values.<br />

Ver<br />

• Outputs the version number of the CPU firmware.<br />