<strong>SonTek</strong>/<strong>YSI</strong><br />
Upon execution, SONREC wakes the <strong>Argonaut</strong> and establishes a direct link to the recorder at the<br />
specified extraction rate. The program then downloads the directory of files stored on the<br />
recorder and displays this information on the screen. The directory screen shows the following<br />
information.<br />
• Number of files on recorder<br />
• List of files including name, size, and date/time of creation.<br />
• Recorder size (bytes)<br />
• Recorder free space (bytes)<br />
• Extraction baud rate<br />
• Active keys and their function, as listed below<br />
Esc - exit program<br />
Spacebar - mark and unmark files for extraction<br />
F3 - retrieve all marked files<br />
Del - format the recorder and erase all data<br />
Within SONREC, move between files in the directory using the up/down arrow and page up/down<br />
keys. The Spacebar will mark and unmark one or more files for retrieval. Pressing F3 retrieves all<br />
currently marked files. The program prompts you to input a destination path for the files; if no<br />
path is given, the current directory is used. While retrieving data files, the program provides<br />
ongoing information about progress with each file.<br />
SONREC transmits data from the recorder in blocks, each of which includes a checksum for<br />
reliable data transmission. If the checksum fails, the program will attempt to transmit the same<br />
block again. If the program encounters a large number of errors during retrieval, it will abort the<br />
process. If this occurs, try running the program again at a lower data extraction rate.<br />
Pressing the Del key will format the recorder and delete all data files. Before the files are deleted,<br />
you are prompted to enter the complete word “Yes” for confirmation. Be certain that all data are<br />
backed up before erasing the recorder. Data cannot be recovered after formatting the recorder.<br />
Pressing Esc will exit the program.<br />
Data files extracted from the <strong>Argonaut</strong> recorder are assigned the extension .ARG.<br />
68<br />