<strong>SonTek</strong>/<strong>YSI</strong><br />

Signal Strength Plot<br />

24<br />

• The signal strength plot shows the three components of signal strength (one for each<br />

receiver) versus time.<br />

• The signal strength vertical-axis is fixed.<br />

• The signal strength time-axis (horizontal) is set to match the velocity plot axis.<br />

• Signal strength data is the best source of diagnostic data for system operation and should<br />

be monitored closely.<br />

• See the <strong>Argonaut</strong> Principles of Operation for details about using and interpreting signal<br />

strength data.<br />

Tabular Display<br />

• The far right portion of the screen shows a tabular display of the most recent <strong>Argonaut</strong><br />

sample.<br />

• Displayed data includes velocity, signal strength, standard error, temperature, and battery<br />

voltage.<br />

• See the <strong>Argonaut</strong> Principles of Operation for details about different types of data.<br />

<strong>Argonaut</strong> Software <strong>Manual</strong> (March 1, 2001)