<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)<br />

<strong>SonTek</strong>/<strong>YSI</strong><br />

3.5. Direct Command Summary<br />

The tables below summarize all direct commands that can be used with the <strong>Argonaut</strong>, including<br />

any abbreviations (shorter versions of the same command). These commands are split into<br />

different sections, and different tables, based on their function. Further details on individual<br />

commands are provided in the sections that follow. The following abbreviations are used for<br />

input parameters to the <strong>Argonaut</strong>.<br />

• d Integer input (e.g., 30)<br />

• d.d Decimal real number input (e.g., 0.33, 1.5)<br />

• yy/mm/dd Date as year, month, and day (e.g., 1996/05/20 or 96/05/20)<br />

• hh:mm:ss 24-hour clock with hour, minute, and second (e.g., 18:15:00)<br />

Help Commands (Section 3.6)<br />

Command Short Function<br />

Help H or ? Available help categories<br />

Help System H/? System General system commands<br />

Help Setup H/? Setup Real-time data collection commands<br />

Help Deploy H/? Deploy Autonomous deployment commands<br />

Help Sdi12 H/? Sdi12 SDI-12 interface control commands (if applicable)<br />

Help Recorder H/? Recorder Data recorder commands<br />

Help Sensor H/? Sensor Peripheral sensor commands<br />

Help Compass H/? Compass Compass/tilt sensor commands<br />

Help Show H/? Show Commands to display system configuration<br />

System Commands (Section 3.7)<br />

Command Short Function Parameters<br />

Start (none)<br />

Starts real-time data collection (enters data<br />

acquisition mode)<br />

Deploy (none)<br />

Starts autonomous deployment (enters<br />

deployment mode)<br />

SaveSetup (none) Save current parameters to EEPROM<br />

Defaults DEF Sets all parameters to factory defaults<br />

Ver (none) Shows CPU firmware version<br />

DSPVer (none) Shows DSP firmware version<br />

BoardRev (none) Shows electronics board revision number<br />

SerNum (none) Shows <strong>Argonaut</strong> serial number<br />

Date (none) Shows / sets system clock date yy/mm/dd<br />

Time (none) Shows / sets system clock time hh:mm:ss<br />

PowerOff (none) Puts <strong>Argonaut</strong> in sleep mode<br />

AutoSleep AS<br />

If ON, <strong>Argonaut</strong> enters reduced power<br />

state during gaps in data collection<br />

On / Off<br />

OutMode OM Selects data output mode Auto / Polled<br />

OutFormat OF Specifies output data format<br />

Ascii, Binary,<br />

or SeaBird<br />

Recorder (none) Turns internal recording on or off On / Off<br />

RecMode (none) Sets recording mode (Buffer not enabled) Normal/Buffer<br />