<strong>SonTek</strong>/<strong>YSI</strong><br />

DSPVer<br />

• Outputs the version number of the DSP firmware.<br />

BoardRev<br />

• Outputs the revision number of the <strong>Argonaut</strong> processing electronics boards.<br />

SerNum<br />

• Returns the instrument serial number from memory (should match the serial number on<br />

the transducer head).<br />

Date [yy/mm/dd]<br />

• Without parameter: returns the date from the <strong>Argonaut</strong> clock.<br />

• When given a date in the form “yy/mm/dd” where yy = year (2 or 4 digits), mm = month<br />

(2 digits), and dd = day (2 digits), it resets the date.<br />

• Example: Date 2001/09/08 or Date 01/09/08 are equivalent commands setting the<br />

date to September 8, 2001.<br />

Time [hh:mm:ss]<br />

• Without parameter: returns the time from the <strong>Argonaut</strong> clock.<br />

• When given a time in the form of “hh:mm:ss” (24-hour clock, where hh = hour (2 digits),<br />

mm = minute (2 digits), and ss = seconds (2 digits)), it resets the time.<br />

• Example: Time 16:24:08 sets the time to 16:24:08 (4:24:08 p.m.).<br />

PowerOff<br />

• Places the <strong>Argonaut</strong> into sleep mode (§3.2.4).<br />

• We recommend placing the <strong>Argonaut</strong> in sleep mode whenever it is not in use.<br />

• The <strong>Argonaut</strong> will automatically enter sleep mode if it is left idle in command mode for<br />

more than five minutes.<br />

AutoSleep or AS [ON|OFF]<br />

• Default parameter: ON<br />

• Without parameter: returns its current setting.<br />

• Determines whether the <strong>Argonaut</strong> enters a reduced power state during data collection.<br />

• When ON, the <strong>Argonaut</strong> will enter the reduced power state between samples. Power<br />

consumption is less than 1 mW in the reduced power state.<br />

• When OFF, the <strong>Argonaut</strong> electronics remain active even when not collecting data.<br />

• Must be OFF to use the Run-Time commands (§3.15).<br />

• In deployment mode, AutoSleep is forced ON.<br />

OutMode or OM [AUTO|POLLED]<br />

• Default parameter: AUTO<br />

• Without parameter: returns its current setting.<br />

• Determines whether data are sent over the serial port after the completion of a sample<br />

(AUTO) or only sent when a specific run-time command is received (POLLED).<br />

• The run-time command “O” (Output) causes the output of the last sample from the buffer<br />

to be sent (§3.15).<br />