CellEnd or CE [d.d] (used in <strong>XR</strong> and SL systems only)<br />

• Default parameter: 2.0 (meters)<br />

• Parameter range: 0.7 to 15.0 (for 1.5-MHz <strong>XR</strong> and SL systems)<br />

0.4 to 6.0 (for 3.0-MHz <strong>XR</strong> and SL systems)<br />

• Without parameter: returns its current setting (in meters).<br />

• When used with a valid parameter, sets the ending location of the measurement volume to<br />

this value in meters. The measurement volume location is specified as the vertical (<strong>XR</strong><br />

systems) or horizontal (SL systems) distance from the transducer head.<br />

• The measurement volume size (the difference between CellBegin and CellEnd) must be<br />

at least 0.5 meters.<br />

• See <strong>Argonaut</strong> Principles of Operation for the effect of sound speed on velocity data.<br />

ProfilingMode or PM [YES|NO] (used in <strong>XR</strong> and SL systems only)<br />

• This command is only available in <strong>XR</strong> and SL systems that have the optional Profiling<br />

Mode feature installed. ProfilingMode allows you to collect a “profile” of data from a<br />

series of range cells. This differs from the standard <strong>Argonaut</strong> method of collecting data<br />

within just one range cell.<br />

• Default parameter: NO<br />

• Without parameter: returns its current setting.<br />

• When used with a valid parameter, enables (YES) or disables (NO) the profiling mode<br />

feature. Setting PM to YES requires you to enter values for the number of cells (Ncells;<br />

NC), cell size (CellSize; CS), and blanking distance (BlankDistance; BD). Data for each<br />

cell is collected and output in the format described in §3.17.<br />

Ncells or NC [d] (used in <strong>XR</strong> and SL systems only)<br />

• This command is only available in <strong>XR</strong> and SL systems that have the optional Profiling<br />

Mode feature installed (see above).<br />

• Default parameter: 1<br />

• Parameter range: 1 to 5<br />

• Without parameter: returns its current setting.<br />

• When used with a valid parameter, sets the number of cells in each profile.<br />

CellSize or CS [d.d] (used in <strong>XR</strong> and SL systems only)<br />

• This command is only available in <strong>XR</strong> and SL systems that have the optional Profiling<br />

Mode feature installed (see above).<br />

• Default parameter: 0.5 meter (for 1.5-MHz <strong>XR</strong> and SL systems)<br />

• Parameter range: 0.25 to 10 meters<br />

• Without parameter: returns its current setting in meters.<br />

• When used with a valid parameter, sets the range cell size in each profile to this value.<br />

BlankDistance or BD [d.d] (used in <strong>XR</strong> and SL systems only)<br />

• This command is only available in <strong>XR</strong> and SL systems that have the optional Profiling<br />

Mode feature installed (see above).<br />

• Default parameter: 0.5 meter (for 1.5-MHz <strong>XR</strong> and SL systems)<br />

20<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)