SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>SonTek</strong>/<strong>YSI</strong><br />
CellEnd or CE [d.d] (used in <strong>XR</strong> and SL systems only)<br />
• Default parameter: 2.0 (meters)<br />
• Parameter range: 0.7 to 15.0 (for 1.5-MHz <strong>XR</strong> and SL systems)<br />
0.4 to 6.0 (for 3.0-MHz <strong>XR</strong> and SL systems)<br />
• Without parameter: returns its current setting (in meters).<br />
• When used with a valid parameter, sets the ending location of the measurement volume to<br />
this value in meters. The measurement volume location is specified as the vertical (<strong>XR</strong><br />
systems) or horizontal (SL systems) distance from the transducer head.<br />
• The measurement volume size (the difference between CellBegin and CellEnd) must be<br />
at least 0.5 meters.<br />
• See <strong>Argonaut</strong> Principles of Operation for the effect of sound speed on velocity data.<br />
ProfilingMode or PM [YES|NO] (used in <strong>XR</strong> and SL systems only)<br />
• This command is only available in <strong>XR</strong> and SL systems that have the optional Profiling<br />
Mode feature installed. ProfilingMode allows you to collect a “profile” of data from a<br />
series of range cells. This differs from the standard <strong>Argonaut</strong> method of collecting data<br />
within just one range cell.<br />
• Default parameter: NO<br />
• Without parameter: returns its current setting.<br />
• When used with a valid parameter, enables (YES) or disables (NO) the profiling mode<br />
feature. Setting PM to YES requires you to enter values for the number of cells (Ncells;<br />
NC), cell size (CellSize; CS), and blanking distance (BlankDistance; BD). Data for each<br />
cell is collected and output in the format described in §3.17.<br />
Ncells or NC [d] (used in <strong>XR</strong> and SL systems only)<br />
• This command is only available in <strong>XR</strong> and SL systems that have the optional Profiling<br />
Mode feature installed (see above).<br />
• Default parameter: 1<br />
• Parameter range: 1 to 5<br />
• Without parameter: returns its current setting.<br />
• When used with a valid parameter, sets the number of cells in each profile.<br />
CellSize or CS [d.d] (used in <strong>XR</strong> and SL systems only)<br />
• This command is only available in <strong>XR</strong> and SL systems that have the optional Profiling<br />
Mode feature installed (see above).<br />
• Default parameter: 0.5 meter (for 1.5-MHz <strong>XR</strong> and SL systems)<br />
• Parameter range: 0.25 to 10 meters<br />
• Without parameter: returns its current setting in meters.<br />
• When used with a valid parameter, sets the range cell size in each profile to this value.<br />
BlankDistance or BD [d.d] (used in <strong>XR</strong> and SL systems only)<br />
• This command is only available in <strong>XR</strong> and SL systems that have the optional Profiling<br />
Mode feature installed (see above).<br />
• Default parameter: 0.5 meter (for 1.5-MHz <strong>XR</strong> and SL systems)<br />
20<br />
<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)