SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>SonTek</strong>/<strong>YSI</strong><br />
Cell Begin / Cell End<br />
• Specify the location of the start and end of the measurement volume. The standard<br />
deviation (accuracy) of the velocity data is inversely proportional to the square root of the<br />
measurement volume size (cell end minus cell begin); larger measurement volume size<br />
gives lower standard deviation. See the <strong>Argonaut</strong> Principles of Operation for more details.<br />
The acceptable range is from 0.5 to 14.8 m.<br />
Data Format<br />
• This parameter is not used with the real-time software.<br />
Dynamic Boundary<br />
• This variable allows the <strong>Argonaut</strong> <strong>XR</strong> to automatically adjust the upper end of the<br />
measurement volume based upon data from the pressure sensor. A setting of DISABLE<br />
turns off the dynamic boundary adjustment and uses the fixed cell location parameters<br />
entered above. A setting of INTEGRATED will use the fixed value of cell begin and will<br />
adjust the end of the cell to be as close as possible to the surface. A setting of LAYERED<br />
will measure a layer whose thickness is defined by (Cell End – Cell Begin), with the upper<br />
edge of the layer as close to the surface as possible.<br />
Wave Spectra<br />
• See Section 8 in the <strong>Argonaut</strong> Operation <strong>Manual</strong>.<br />
Record File Name<br />
• This specifies a root file name for data files recorded by the real-time software.<br />
• Up to five letters and/or numbers can be specified.<br />
• A 3-digit termination is added to the specified file name, starting with 001 and<br />
incrementing (002, 003, …) each time recording is stopped and started.<br />
• The file names use the extension .ARG. For example, if the file name is TEST, the first file<br />
will be TEST001.ARG, followed by TEST002,ARG, etc.<br />
• Files are stored in the directory shown under the File|Open option.<br />
Comment 1/2/3<br />
• Up to three comment lines are provided for you to document the data set.<br />
<strong>Argonaut</strong> Software <strong>Manual</strong> (March 1, 2001) 21