01.06.2013 Views

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>SonTek</strong>/<strong>YSI</strong><br />

Upon completion of the measurement, a service request “a” is sent to the data recorder<br />

indicating the sensor data is ready. The recorder may wake the sensor with an SDI-12 BREAK<br />

(continuous spacing by the data recorder on the data line for at least 12 ms) and collect the data<br />

anytime after the service request is received, or when the specified processing time has elapsed.<br />

Examples of “aMc!” commands:<br />

Command Response Time Values Description<br />

“aM!” “a0609” 60 s 9 Make one measurement from CB through CE.<br />

“aM3!” “a0004” 0 s 4 Make a measurement of Cell 3 (if ProfilingMode=Yes).<br />

Note: Report data taken at the time of the last “aM!”<br />

command over the range set by the BC, CS, and NC<br />

commands (see Note †† on page 102 regarding “aMn!”).<br />

104<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)