A4.4. Configuring the <strong>Argonaut</strong>-SL/<strong>XR</strong> for use with the AO Converter<br />
<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)<br />
<strong>SonTek</strong>/<strong>YSI</strong><br />
To configure the <strong>Argonaut</strong> SL or <strong>XR</strong> for operation with the AO Converter, use the following<br />
command sequence. We assume you are using SonTerm to “talk” to the Converter.<br />
1. Choose the AO type (DISABLED, CURRENT, or VOLTAGE).<br />
AOT CURRENT (sets AO to use current-scaling)<br />
2. Set desired data output format. Note that this entry will clear all AO channels. Do NOT<br />
change this value after you have set the AO channels; otherwise, you will have to re-enter all<br />
channel settings using the SAO command.<br />
OF ENGLISH (tells <strong>Argonaut</strong> to use English measuring units)<br />
3. Configure all the AO settings for your application.<br />
SAO 1 VELMAG –3 3<br />
(sets <strong>Argonaut</strong>’s AO channel 1 to send velocity magnitude to address 1 of the Converter<br />
using minimum and maximum values of 3 ft/s)<br />
4. Configure the <strong>Argonaut</strong>’s regular settings such as CB, CE, AI, SI, etc. (see the <strong>Argonaut</strong><br />
Operation <strong>Manual</strong>).<br />
5. Save your settings.<br />
SaveSetup<br />
6. Start data collection.<br />
Start<br />
7. Verify that the <strong>Argonaut</strong> outputs the AO data string, where 2.83 is the current value in mA.<br />
This value will vary depending on the flow conditions.<br />
$1AO+00002.83<br />
8. Connect the male, DB-9 RS232 connector on the AO splitter cable (Figure A4.2) to the<br />
female, DB-9 RS232 connector on the <strong>Argonaut</strong> power and communications cable.<br />
9. The <strong>Argonaut</strong>-AO system is now ready for data collection.<br />
Notes:<br />
a. If recording is enabled, the analog data will be stored on the recorder in addition to being<br />
sent to the COM port as an analog output.<br />
b. If AOT is set to DISABLED, the data will be sent to the serial port according to the OutFormat<br />
setting.<br />
c. If the OF command is sent to change the output format (English, Metric) after entering the<br />
AO commands, this will reset all AO settings. You will then have to re-enter all AO settings<br />
using the SAO command.<br />
117