• Parameter range: 0.4 to 10 meters<br />

• Without parameter: returns its current setting in meters.<br />

• When used with a valid parameter, sets the blanking distance to this value.<br />

<strong>Argonaut</strong> Operation <strong>Manual</strong> Firmware Version 7.9 (May 1, 2001)<br />

<strong>SonTek</strong>/<strong>YSI</strong><br />

DynBoundaryAdj or DBA [YES|NO] (used in <strong>XR</strong> systems only)<br />

• This command is only available in <strong>XR</strong> systems.<br />

• Default parameter: NO<br />

• Without parameter: returns its current setting.<br />

• When set to YES, the <strong>XR</strong> dynamically adjusts the values of CellBegin and CellEnd to<br />

reflect the changes in a changing water level. Examples: (1) To monitor the currents in the<br />

upper 2 m of water below the surface, set CB to 13, CE to 15, and DBA to YES; (2) To measure<br />

currents throughout the entire water column, set CB to 0.4, CE to 15, and DBA to YES.<br />

• When set to NO, the <strong>XR</strong> uses the fixed cell location parameters (CellBegin and CellEnd)<br />

entered above. Example: When monitoring currents in a fixed layer (1 m above the<br />

bottom) set CB to 1, CE to 2, and DBA to NO.<br />