01.06.2013 Views

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SonTek/YSI Argonaut-XR Technical Manual - HydroScientific West

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>SonTek</strong>/<strong>YSI</strong><br />

5.5. Windows Software Support<br />

<strong>SonTek</strong> Windows-based software offers an alternative to the DOS extraction programs for handling<br />

the external pressure sensor calibration data.<br />

5.5.1. Downloading Calibration File Using SonTermW<br />

Starting with v2.0, SonTerm for Windows (SonTermW) provides an easy way to load the frequency<br />

pressure sensor calibration file into the system processor. Simply establish communication<br />

with the system in SonTermW, click the Load Druck Calibration icon, choose the appropriate<br />

file, and the load process will be performed.<br />

5.5.2. Specifying Calibration File When Using ViewHydra<br />

When opening a data file (.adr) with ViewHydra, you must supply a sensor calibration file (.drk)<br />

to convert pressure data to dbars. If no pressure calibration is provided, all pressure data will be<br />

displayed in raw frequency counts. ViewHydra lets you save the calibration data in an associated<br />

data file workspace, so the calibration data is loaded automatically the next time the data file is<br />

opened.<br />

After the data file is loaded, you can select the sensor offset from the processing menu to convert<br />

the absolute pressure reading.<br />

8<br />

1. Run DRUCK/PAROSFREQ in SonTermW and make note of the pressure reading with the system<br />

in the air (say 10.025 dbar for this example).<br />

2. In the Processing|Sensor Calibration Settings… menu, set Sensor Type to Ext Pressure, and<br />

then enter a value of –10.025 (for this example) in the Offset field to convert pressure data<br />

into strain gage data (relative to the atmosphere).<br />

3. Now use the Processing|Recompute Burst Statistics File… option to propagate the changes<br />

into the burst-averaged data.<br />

4. When exiting the program, use the Save (Overwrite) Workspace option to store the calibration<br />

and offset data. The program will now remember this information for this data set.<br />

5.5.3. Using ViewADP with PCADP Data and Waves<br />

Data collected by a standard ADP stores mean pressure data in dbars; no additional conversion is<br />

needed. However, PC-ADP data and ADP wave series data (P-SERIES & PUV-Series) require<br />

pressure calibration.<br />

When opening a data file with ViewADP, click the Load ExtPres CalFile button to specify the<br />

calibration file.<br />

Notes:<br />

1. If pressure calibration is not loaded, all pressure data will be displayed as raw frequencies.<br />

2. If pressure calibration data is not loaded, wave data processing is disabled.<br />

3. ViewADP will save the calibration data in the associated data file’s workspace. This allows<br />

the calibration data to load automatically the next time the data file is opened.<br />

Using Frequency Pressure Sensors with <strong>SonTek</strong>/<strong>YSI</strong> Systems (June 2001)

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!